11
Technical data
Order number iba
10.120300
Mechanical stability and
test parameters (all 3 axes)
DIN IEC 68-2-6;
2 g rms 90 Min @ 0..250 Hz / axis (all ibaPADU-8 types);
EMI test parameters
EN 55011 (Class A); EN61000-4-6 (Class 3); EN61000-4-3/ENV
50204 (Class 3)
Operating temperature
0
°
C to 50
°
C (32 °F…122 °F),
other ranges on request, e.g. –20 °C ....+50 °C (-4 °F...122 °F)
Storage temperature
-25
°
C to 70
°
C (-13 °F…158 °F)
Transport temperature
-25
°
C to 70
°
C (-13 °F…158 °F)
Cooling
Self cooling
Mounting
On DIN-rail, snap-on
Humidity Class
F no moisture
Protection Class
IP20
Power supply
24 V DC +/-20 % non stabilized
Power/Current consumption (w/o
load)
typ. 300 mA, max. 400 mA; I
OFF->ON
approx. 1A
Sampling rate programmable up to 25 kHz 8 channels simultaneously
FO-cable
Coupling
62.5/125 µm
ST Lean
Max. length of FO between devices 2000 m (6560 ft)
Dimensions (WxHxD) [mm] ([inch]) 54 x 194 x 155 (2.13 x 7.64 x 6.10)
Weight (incl. packaging and docu-
mentation)
1050 g (37 oz)
Analog inputs
Number
8
Each channel has its own
ADC
Resolution
14 Bit
Input level / type
Ri
ON
Ri
OFF
+/-10 V
100 kOhm @100 Hz;
10 kOhm @ 5 kHz
50 kOhm
Keep signal lines as short as
possible
Sampling rate
Typ. 25 kHz
8 channels simultaneously
Frequency range
0…25,000 Hz
Dynamic range
84 dB
RC Low pass
12.5 kHz / 3 dB, 25 kHz / 6 dB (fixed)
Galvanic Isolation
Channel / Channel/ Device-ground
1.5 kV
Binary inputs
Number
8
Input level
log 0; < +/-10 V
log 1; > +/-10 V
Sampling
max. 25 kHz
Simultaneous sampling with
analog inputs
Galvanic isolation
Channel/Channel/Digital Ground
1.5 kV
1.5 kV