ELECTROMAGNETIC INTERFERENCES 

 
The I-TECH LA500 equipment has been designed and manufactured according to the 
ELECTROMAGNETIC  COMPATIBILITY  DIRECTIVE  2004/108/CE  with  the  aim  of 
providing adequate protection from harmful interference when installed in homes 
and health establishments. 
 
The I-TECH.AR equipment does not generate significant radio frequency energy and is 
adequately  immune  to  radiated  electromagnetic  fields.  Therefore  it  does  not 
detrimentally  interfere  with  radio-electric  communications,  electro-medical 
equipment  for  monitoring,  diagnosis,  therapy  and  surgery,  office  electronic  devices 
such as computers, printers, photocopiers, fax machines, etc. or 
any  electric  or  electronic  equipment  used  in  these  environments,  as  long  as  said 
equipment complies with the ELECTROMAGNETIC COMPATIBILITY directive. 
 
In  any  case,  in  order  to  avoid  any  interference  problems,  we  recommend  that  you 
operate  the  therapy  equipment  far  enough  away  from  critical  equipment  for 
monitoring vital patient functions, and that you be careful when applying therapy to 
patients with pacemakers. 
 
 
 
 

TROUBLESHOOTING CHART 

PROBLEM 

POSSIBLE CAUSE 

SOLUTION 

The LCD display in 

the frontal panel 

does not turn on: 

the device does 

not work. 

The power plug is inserted not 

correctly in the socket. 

Check that the socket works properly. 

The power cable is not inserted 

correctly into the connector of the 

device. 

Insert correctly the plug and the cable 

in the connector of the device. 

The power cable is worn out or 

interrupted 

Replace the power cable. 

The emergency switch is off. 

Turn on the emegency switch. 

Fuse/s is/are defective or 

interrupted. 

Replace missing, defective or 

interrupted fuse/s 

The control electric circuit is 

broken. 

Contact IACER srl service centre. 

The LCD display in 

the frontal panel 

does not turn on. 

Defective components in the 

control electric board. 

Contact IACER srl service centre. 

Some controls on 

the front control 

panel do not work 

properly. 

Faulty keys or buttons. 

Contact IACER srl service centre. 

Control electronic circuit failure. 

 

The device does 

not emit the 

therapy. 

Parameters 

are 

not 

setted 

correctly. 

Check that the parameters have been 
setted correctly. 

Laser  sources  not  working  or 
depleted. 

Check laser source emission mode. 

Faulty  components  in  the  control 
electronic circuit. 

Contact IACER srl assistance centre. 

 

Faulty supply of laser sources. 

 
 

The unit works 

properly but with 

a significant 

decrease in 

efficiency of the 

treatment. 

Faulty or depleted laser source. 

Contact IACER srl assistance centre. 

Possible break down in power 

generator circuit of the unit. 

The equipment 

starts up, or 

seems to work 

properly, but 

there is no 

emission. 

No safety key or the interlock 

circuit is open. 

Insert the DIN safety key into the 

front socket. Reset the safety 

conditions. 

 

 

Содержание LA500

Страница 1: ...I A C E R Srl MNPG127 02 06 06 17 LASER THERAPY...

Страница 2: ......

Страница 3: ...E 8 INDICATIONS 8 CONTRA INDICATIONS 8 DEVICE DESCRIPTION 9 PATIENTS PREPARATION 9 HOW TO USE IT 9 Therapeutic indications 11 MAINTENANCE 13 HANDLES CLEANING 13 WORKING PROBLEM 13 ELECTROMAGNETIC INTE...

Страница 4: ...important information and useful tips for operating the equipment CAUTIONS The CAUTION message appears before operations which if not correctly performed may cause damage to the machine and or its acc...

Страница 5: ...t use the equipment in places where it might get wet Before operating the machine carefully check the correctness of the connections according to the instructions Do not use accessories other than the...

Страница 6: ...ecution of different procedures from those specified in this user manual can cause the exposition to dangerous radiation The operator has the responsibility to verify that the issuing head remains wel...

Страница 7: ...er production of ATP The applications of the I TECH LA500 laser produce a number of effects on the treated tissues 1 increase in hematic flow vasodilatation of capillaries and arteries 2 biostimulatio...

Страница 8: ...ensitive to laser radiation and can be permanently damaged from direct or reflected laser beams The special safety glasses must be worn by both patient by the operator pregnancy the laser is contraind...

Страница 9: ...of the patient for the therapy carry out the following operations 1 wash the skin using soap and mild soap or alcohol where the head of the laser will be positioned 2 dry well the skin HOW TO USE IT...

Страница 10: ...ent time last 3 minutes Select the operation mode CONTINUOS or CYCLE using the buttons on the right side of the front panel picture below With CYCLE mode it is possible to select the ON time from 1 to...

Страница 11: ...ition upper picture Take away the key from the machine to avoid the use by unauthorized personnel Please do not switch off the device unplugging the cable from the power socket to avoid any type of da...

Страница 12: ...ITIS 0 3 66 sec 264 sec BAKER S CYST 0 3 66 sec 264 sec DE QUERVAIN SYNDROME 0 2 100 sec 400 sec OEDEMA 0 3 66 sec 264 sec HEMATOMA 0 3 66 sec 264 sec EPICONDYLITIS 0 4 50 sec 200 sec HERNIATED DISC 0...

Страница 13: ...ent HANDLES CLEANING The laser handle is very weak and it needs a daily cleaning The following advices are absolutely important to not damage the lens and the laser probe It is suggested to 1 Remove t...

Страница 14: ...that the socket works properly The power cable is not inserted correctly into the connector of the device Insert correctly the plug and the cable in the connector of the device The power cable is wor...

Страница 15: ...B Output channels 1 Class of isolation parts applied according to the rule EN 60601 1 I BF Protection level against the liquids penetrations according to the rule EN 60601 1 IPX0 Command of execution...

Страница 16: ...that the product shall not be disposed of as domestic waste The user must dispose of scrap equipment by taking it to a recognized centre of recycling of electrical and electronic equipment Appendix B...

Страница 17: ...ctrostatic discharge ESD IEC 61000 4 2 6kV contact 6kV contact Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30...

Страница 18: ...th accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the ME EQ...

Страница 19: ......

Страница 20: ......

Страница 21: ...nd warehouse 30030 MARTELLAGO VE Via S Pertini 24 A Tel 0039 041 5401356 Fax 0039 041 5402684 Cod Fisc P IVA IT 00185480274 R E A VE N 120250 M VE001767 e mail iacer iacer it Internet www itechmedical...