MNPG127-02
Pag. 15
TECHNICAL FEATURES
Power supply:
230 Vac, 50-60 Hz, ±10%
Maximum absorbed power
20 VA
Double protection fuse on power
supply (T):
1,5 A-T - 5 X 20 mm
LCD display
Icons
Maximum Power
500 mW ± 20%
Laser diode wavelength
810 nm
Laser classification
3B
DNRO (m)
2,3
Divergence
260 mrad
Impulse duration
set
Emission
CONTINUOUS
Adjustable percentage of emitted power
0% - 100%
Emission frequency
set
Pulsed mode
ON/OFF (sec)
Pointing led
Red light
Classification in compliance with the directive 93/42/CEE
IIB
Output channels
1
Class of isolation / parts applied according to the rule EN 60601-1
I / BF
Protection level against the liquids penetrations according to the
rule EN 60601-1
IPX0
Command of execution of the treatment
Button
Configuration of the laser probe
Diode probe
Defocused
Connection specific for the device
Adjustable treatment time
0-30 minutes
Diameter of the spot on the skin of
the patient
< 10 mm
Dimensions (Length * Height * Depth) :
30x15x12 cm
Weight main body:
5,2 Kg
Conditions of use
temperature environment
(+5 : +40) ° C
relative humidity
< 80 % without condensation
Storage / transport conditions
temperature environment
(+5 : +40) ° C
relative humidity
<93% without condensation
Atmospheric pressure
(500 : 1060) hPa