MNPG127-02
Pag. 11
Press the WAIT button to confirm all the adjusted values: at this time it will be not
possible to modify the treatment time and the power. Press the START button to start
the treatment.
When the treatment will be finished, the machine will emit an intermittent signal. If
you need to stop the treatment before the end press the STOP button and the device
will stop immediately the laser emission.
In case of an emergency stop press the red button placed on the back of the machine.
After the treatment switch off the device moving the key switch on the OFF position
(upper picture).
Take away the key from the machine to avoid the use by unauthorized personnel.
Please do not switch off the device unplugging the cable from the power socket to
avoid any type of damage to the device.
Therapeutic indications
The following table shows the suggested values for the main therapeutic
applications considering 20 J/cm
2
and with a continuous emission)
The same values can be applied with a pulsed emission (by changing the duty
cycle) with longer treatment time.
The given indications can be changed depending of the discretion of the therapist.
PATHOLOGY
POWER Watt
TIME for each
point (1
cm
2
)
TOTAL TIME
(4 trigger points )
PAIN RELIEF (acute pain)
0,1
200 sec
800 sec
PAIN RELIEF (chronic pain)
0,5
40 sec
160 sec
NEK PAIN (acute pain)
0,1
200 sec
800 sec