MNPG127-02
Pag. 10
Connect the interlock to the plug placed on the back of the machine.
Connect the probe laser cable to the proper plug placed on the front of the machine.
STARTING
After the machine has been placed and installed using the instructions given in the
previous chapters, plug the power cable into the wall socket (230 Vac) and switch on
the device moving the key button on the “ON” position witch is placed on the front
of the machine.
This operation prepares I-TECH LA500 for use, determining the switching on of the
LCD display that shows that the equipment is ready to operate. The default program
is pre-setted on CONTINUOS mode and the treatment time last 3 minutes.
Select the operation mode (CONTINUOS or CYCLE) using the buttons on the right side
of the front panel (picture below).
With CYCLE mode it is possible to select the ON time (from 1 to 9 seconds) and the
OFF time (from 1 to 9 seconds) using the buttons on the right side of the front panel
(next two pictures ).
Select the treatment time using the TIME + and TIME – buttons placed on the left
side of the front panel of the machine (picture below). You can select the treatment
time from 30 seconds to 30 minutes.
You can set treatment intensity with the POWER + e POWER – buttons placed on the
left side of the front panel. The default power is setted to 200mW and the range is
adjustable from 10mW to 500mW in according to treatment needs.
After the set-up of the power place the laser probe on the area to be treated and
fasten it with the elastic belt.