Error messages / Troubleshooting
FluidAqua Mobil - FAM10/15-P
en(us)
Page 103 / 124
BEWA FAM10-15-P 4052078 en-us 2015-11-05.doc
2015-11-05
1. Message:
26. Heater safety thermostat triggered.
2.
Check whether the safety thermostat has tripped.
The factory setting is 80°C.
3.
->a. The safety thermostat has
triggered.
-> proceed to Step 4.
->b. The safety thermostat has not
triggered.
-> proceed to Step 10.
4.
Reset a safety thermostat.
See page 102 for details.
5. ->a.
Safety
thermostat
reset.
-> proceed to Step 7.
->b. Safety thermostat reset not yet
possible after 3 hours.
-> proceed to Step 6.
6.
Replace the safety thermostat
on the heater.
-> proceed to Step 18.
7.
To reset the fault, press the F3 button.
8.
->a. Error reset
-> proceed to Step 21.
9.
->b. Resetting fault not possible.
-> proceed to Step 10.
10.
Switch off the FAM at the main switch. Pull power plug.
11.
Check the safety thermometer cable up to the PLC.
->a. The cable is defective.
-> proceed to Step 17.
12.
->b. The cable is not defective.
-> proceed to Step 13.
13.
Replace the K7 relay.
14.
Plug the power plug back in. Switch on the FAM at the main
switch and start operation by pressing the F1 key.
->a. The fault is repeated.
-> proceed to Step 16.
15.
->b. The fault is not repeated.
-> proceed to Step 21.
16.
The PLC is faulty.
-> proceed to Step 20.
17.
Replace the cable and/or plug.
18.
Plug the power plug back in. Switch on the FAM at the main
switch and start operation by pressing the F1 key.
->a. The fault is repeated.
-> proceed to Step 13.
19.
->b. The fault is not repeated.
-> proceed to Step 21.
20.
Contact the HYDAC service department.
21.
The fault is repaired.