6 | Operation
HYDAC FILTER SYSTEMS GMBH
58 / 124
BeWa CTU1x3x-CTU1x4x 4093420f en-us web
Fig. 24:
Analysis chamber with basket mount
The dimensions of the hook-in basket are 220 x 110 x 40 mm.
Mesh width = 4 mm.
Proceed as follows for spray extraction:
ü
Determine the negative control value; for details, see
mining the negative control value - Settings/parameters
[
Flush the inner chamber including the auxiliary equipment
such as the hook-in basket, grating, mounts for the test item,
etc.
1. Place the test item to be sampled in the analysis chamber.
2. Close the door and lock it.
3. Select the required analysis parameters on the user interface
and start the spray extraction.
4. Spray the component surfaces to be analyzed evenly.
The test liquid (now referred to as the analysis fluid), loaded
with the partially rinsed off particulate contamination, is
drawn across the filter membrane using a vacuum.
ð
Wait until the analysis fluid has been fully extracted through
the filter membrane.
1
2
3
4
5
6
7
8
9
A