7 | Carrying out the maintenance / inspection
HYDAC FILTER SYSTEMS GMBH
88 / 124
BeWa CTU1x3x-CTU1x4x 4093420f en-us web
For cleaning the analysis chamber, we recommend a moist, clean,
commercial microfiber cloth. Finish cleaning the analysis chamber
with a long Flushing of inner chamber.
7.2.1 Cleaning the analysis chamber in the case of test
liquid - Water with surfactants
Clean the analysis chamber when using a water-based test liquid,
due to self-contamination - daily.
7.2.2 Cleaning the analysis chamber in the case of test
liquid - G60 Spezial/Feroclean L261
Clean the analysis chamber if using test liquid G60 Spezial /
Feroclean L261 - weekly.
7.3 Cleaning the door
Clean the door weekly with a moist, lint-free cloth.