Operating and Maintenance Instructions : FCU 2110-5 / FCU 2210-5
Edition 1.10
HYDAC Filtertechnik GmbH
Page 43 of 52
7.4 SETUP Menu
Purpose
This menu enables the user to enter basic settings or change them. The settings marked in
the figure below are preset in the factory and should not be subsequently changed by the
user.
Operation
Battery state
Autostart
Abort at Q = 0
Date / time
Busadress
Operating hours
Pump pre-run time
SETUP-key
Autostart:
Abort at Q=0:
(mode M1, M5 only)
→
yes
If the current flowing through the sensor drops to “0”,
measurement is discontinued and the FCU switched
off.
→
no
If the current flowing through the sensor drops to “0”,
measurement is discontinued, however the FCU
remains switched on and resumes measurement as
soon as the flow rate is sufficient again.
Pump pre-run time
When conducting measurements with the aid of the KE1310
conditioning unit or when using an external filter pump unit (e.g.
OF5C), this function enables the user to operate the pump until
the pressure and flow rate are stabilized and the FCU is to
resume measurement. The setting time may range from 1 to
200sec.
Recommended: 60 sec.
A setting of 0 sec disables this function.