6.2 Hydraulic Installation and Commissioning
6.2.1 Measurements with Suction INLET
Figure 5 : Setup for measurements with suction inlet
After measurements with high pressure inlet you should wait some
minutes before switching to position SUCTION to be sure that the
internal accumulator has reliefed its pressure to the Outlet
A
Plug the return hose (DN 7) onto the "Outlet" connector on the back panel and route it into
a suitable container, e.g. hydraulic tank or any other container.
B
Switch the pump switch on the rear panel to ON position (refer to Figure 3 : Rear View) and
turn the flow control valve on the front panel clockwise to end position .
C
Set the ball valve on the rear of the FCU to position SUCTION (refer to Figure 6 : Ball Valve
Position "Suction").