FluidControl Unit - FCU 1000 Series
HYDAC Filtertechnik GmbH
en
Page 54
BeWa_FCU1000_3371346a_en.doc
2007-12-04
14 Technical
Data
Contamination Sensor
Self-diagnosis
continuously with error indication via status LED and
display
Display
LED, 6 / 4 / 4 digits, in 17 segment format
Measured values contamination ISO code / SAE class / NAS class
Measured values temperature
-25 ... 100°C / -13 ... 212°F
Measured values saturation
0 ... 100 %
Measurement range
Display ISO classes min. 9/8/7 ... max. ISO 25/24/23
Calibrated in the range ISO 13/11/10 - ISO 23/21/18
Accuracy
± 1/2 ISO-Code
Service Display
Flow / Drive
Measuring time programmable 10 ... 300 seconds
Hydraulic data
Suitable Fluids
Mineral oil
Hydraulic Connectors
INLET:
Test connector type 1604
OUTLET:
DN7 nipple socket
INLET operating pressure
-0.5 ... 45 bar / 0 ... 650 psi
With high pressure adaptor
15 ... 345 bar / 217 ... 5000 psi
OUTLET operating pressure
0 ... 0.5 bar max. / 0 ... 7.5 psi
Measurement flow rate:
30 ... 300 ml/min (viscosity dependant)
Permissible viscosity range
10 ... 350 mm²/s / 46 ... 1622 Sus
(for hydraulic oil up to ISO VG 68)
Maximal suction height
1 m
INLET suction hose
DN4, open-end, color: clear-transparent, Length = 0.3 m
INLET pressure hose
DN4 with screw connection for test point 1620,
color: black, Length = 2 m (78.74 inch)
OUTLET return line hose
DN7, open-end, color: clear-transparent, Length = 1 m