CTM characteristics
CTM characteristics
The CTM-SC is a specially designed supply module which only makes up an
extraction unit to determine the surface cleanliness of components when
used in combination with a corresponding CTM-E extraction module.
Functional principle
After entering the required analysis parameters in the CTM-SC software, you
push the foot switch to pump the test liquid from a reservoir, via a purification
filter, to one of the three valves with the corresponding connection (
A / B / C
)
on to the extraction module CTM-E.
The test liquid, (now called the analysis liquid) carrying the particles washed
off of the component, is sucked through the analysis membrane using a
diaphragm pump and flows back through the connection (
R
) into the
reservoir.
The level in the reservoir is constantly monitored. If the fill level of the
reservoir drops to the lower limit, the extraction process will be interrupted.
The set test liquid volume is measured using a volume flow meter. Once the
preset amount is reached, the intake is stopped.
CTM-SC 3xxx series
en(us)
Page 18 / 56
BeWa CTM-SC 4165698b en-us 2017-05-10.docx
2017-05-10
Содержание CTM-SC 3 Series
Страница 55: ......