8
vi. Fill
ON
, Flow
OFF
, Sump
ON
1. System will begin filling water and pumping water out at the
same time.
WARNING:
In this case the sump pump will not turn off
and will continue to pump out water if the water level gets below the
lower water level sensor.
2. This setting should not be used for anything other than recycling all
water out of the tank and refilling with fresh water. However, the
user will need to manually turn off the sump pump.
vii. Fill
ON
, Flow
ON
, Sump
OFF
1. System will flow water through the machine and will fill the water
tank until water is detected at the upper water level sensor. Once
water is detected the filling will stop but water flow will continue.
viii. Fill
ON
, Flow
ON
, Sump
ON
(Recommended setting during testing).
1. System will fully automate water cycling. Water will flow over the
sample continuously. At the same time, water will be pumped out
of the tank until no longer detected on the lower water level sensor.
At this point the sump will turn off and the water tank will fill up until
water is detected at the upper level sensor. The sump will then
re-activate, starting the cycle anew.