26
Logger ID
Each machine that is connected to your network requires a unique Logger
ID. These numbers can be assigned any number between 1-245. Only one
machine can be assigned a specific ID, If there are other Humboldt testing
devices on the same Local Access Network, please ensure that each has
its own unique ID number.
Sound
Checking this box enables or disables audio feedback for
logger events. If the box remains unchecked, the machine will
produce an audible beep each time a button is pressed.
Push Speed
The rate the sample will advance when a test is run under
continual advance mode (see Test Controls on page 11).
Run Speed
This determines the water flow rate during a test. This value
can also be changed on the Home page when preparing to
run a test (see page 10, item 2)
Power Rating
This setting is dependent on the pump motor type installed
and will not typically need to be adjusted. The default value is
set at 15 hp.
Orientation
This is dependent on the version of the EFA produced. This
will be set at the factory when the pump is initially installed
and should not need to be adjusted by the end user.
Limit Switch
The limit switch prevents the sample extrusion piston from
overextending and damaging the machine once the sample
extrusion piston has reached its maximum extension. It is
recommended that this setting remain in the factory default
of ENABLED.
Содержание HM-5940
Страница 1: ...product manual 4 22 HM 5940 Erosion Function Apparatus ...
Страница 5: ...5 Figure 2 Elevation View of EFA Figure 3 Plan View of EFA ...
Страница 6: ...6 Figure 4 Detail View of Sample Compartment and Function of Components Within ...
Страница 25: ...25 This screen is accessed by clicking on the Preferences button 1 1 ...