23
You should see this screen:
Press
(SELECT CHANNEL) to scroll through and choose the
channel you wish to calibrate. Once you have chosen the desired
channel you wish to calibrate press
(CLEAR CALIBRATION).
WARNING: IF YOU CLEAR A CALIBRATION IT IS LOST FOREVER!
After the calibration has been cleared, Press
(CALIBRATE
CHANNEL) You should see this screen:
Press
(SET GAIN). With the instrument you wish to calibrate
mounted to a calibration frame attached to the HM-3000, record the
division min value with no load on the instrument (X divisions).
Next, apply the full engineering load value to instrument on the
calibration frame and again, record the division min reading (X
divisions). Subtract the two values. This value should be a number larger
than the maximum rating of the instrument.
If not, you need to increase the gain to compensate for the difference.
Press
(SET GAIN) and repeat until the difference between the two
values is more than the maximum rating of the instrument.
CALIBRATION SCREEN
* SELECT CHANNEL “CH1`”
* CALIBRATE CHANNEL
* CLEAR CALIBRATION
“CH1” CALIBRATION
* SET GAIN
* SET MIN
* SET MAX