Figure 5. HM-2750 Series Front Control Panel
Key Functions
13
Set Up Key
Pushing the Set Up Key brings up the main set up screen
from which you can run a test, review your last test, access
the engineering set up menus and set the date, time,
units, standards criteria. This key also is used to return to a
previous screen.
Speed Key
Pushing the Rate Key allows you to change the selected
platen speed between 0.00001 and 0.499999 inch/min.
Stop Key
Pushing the Stop Key stops the machine.
Forward Key
Pushing the Forward Key moves the shear box to the right.
Also, by holding down the Key, the machine will move
forward at the fast rate of 0.5 in/min. Releasing the key
changes the machine's speed back to its set rate.
Reverse Key
Pushing the Reverse Key moves the shear box to the left.
Also, by holding down the Key, the machine will move in
reverse at the fast rate of 0.5 in/min. Releasing the key
changes the machine's speed back to its set rate.
SPEED
STOP