20
Set Min
The Engineering Zero (ENG MIN) is not configurable and is set at 0.
Set Max
Press
(ENG MAX) to set the Engineering Maximum value.
You should see this screen:
Press
(ENG MAX) to move between the numbers. Keep pressing
until the digit you want to change is flashing. Once you have
selected the number you want to change, press
(INCREASE) or
(DECREASE) to change the number to the desired value. Below
is a configuration chart for Humboldt instrumentation. These are the
numbers that correspond to each Humboldt device.
Load Cell
Part#
Eng Max (lbf) Metric Max (kN)
10,000 lb. (50kN)
HM-2300.100
010000
050.00
5,000 lb. (25kN)
HM-2300.050
005000
025.00
2,000 lb. (10kN)
HM-2300.020
2000.0
10.000
1,000 lb. (5kN)
HM-2300.010
1000.0
05.000
500 lb. (2.5kN)
HM-2300.005
0500.0
02.500
LVDT Transducer
Part#
Eng Max (in.) Metric Max (mm)
0.4" (10.0mm)
HM-2310.04
0.4000
10.000
1.0" (25.4mm)
HM-2310.10
01.000
25.000
2.0" (50.8mm)
HM-2310.20
02.000
50.000
Pore Pressure
Part#
Eng Max (psi) Metric Max (kPa)
100psi (1000kPa)
HM-4170
0100.0
001000
To return to any prior screen you can always press
“CH1” MAX SETUP
* ENG MAX =
“000000”
* INCREASE
* DECREASE