www.humancaregroup.com
33
ENGLISH
Care instructions for the lift and accessories
Care instructions for the lift
• The lift should be recharged every night
• Wipe the ceiling lift with a slightly dam-
pened cloth with a water solution using a
mild detergent.
• Never immerse the lift in water, or use any
other fluids to clean the lift.
• NEVER use strong detergents on the ceiling
lift.
• If the lift needs to be disinfected, Du-
PontTM Virkon® can be used
• The lift belt must be washed and disinfec-
ted as instructed above
• Activate the emergency stop if the ceiling
lift is not to be used for a few days, oterwi-
se the batteries will run down
• The lift must always be stored or moved
lying horizontally
Annual inspection
The lift must be inspected at least once a year
by someone with Human Care qualifications
and in accordance with Human Care’s directive.
Pay particular attention to parts that show wear.
The lift’s service indicator will flash yellow when
a service needs to be performed.
Daily inspection
• Always check the sling before use. There
must be no loose seams or other signs of
wear to lift belts, belts and fabric. If there
are, replace the sling
• Check the ceiling lift for visible signs of
damage
• Check that the lift is correctly mounted in
the distance strap or rail roller
Monthly inspection
Check the lift belts for damage or wear and tear.
Feed out enough of the lift belts to allow their
entire length to be checked. Contact a retailer if
the belt is damaged.
Service
Installation, service and maintenance must only
be performed by personnel with Human Care
qualifications and in accordance with Human
Care’s directive. Only original spare parts may
be used.
Service agreements
Human Care invites you to sign a service agre-
ement for annual inspection and test loading.
Transport and storage
During transport, or when the lift is not going to
be used for some time, the mains switch must
be off or the emergency stop button pushed
in, otherwise the batteries will self-discharge.
The lift must be transported and stored lying
horizontally. The climatic conditions should be as
follows: ambient temperature from 0 °C to 40 °C,
relative ambient humidity from 30% to 80% and
ambient pressure from 790 hPa to 1060 hPa.
Product lifetime
The product´s expected lifetime is ten (10)
years or 10.000 lifts, if the product is used as
intended and maintained according to the
manufacturer´s instructions, depending on the
intensity of use and maximum load applied
during use. If the product label is no longer
legible, the product should be discarded.
Warranty and support
If you need information or support, please con-
tact your Human Care office, www.humancare-
group.com, or distributor.