Setup mode
Chapter 4
OPERATION MANUAL
V2.2.0en/09.08.21//17.12
Unichiller® eo Desktop
57
4.2.12
Setting the setpoint thresholds
The limits for the minimum and maximum setpoint serve the safety of your system. They
must
be
set for the application range of the thermal fluid before starting the first temperature control and
when changing the thermal fluid. The maximum setpoint limit limits the setpoint setting for the bath
or flow temperature. The minimum setpoint limit protects against high viscosity or freezing at low
temperatures. The adjustable setpoint is then only available in the temperature range between the
minimum and maximum setpoint limit.
PROCEDURE
Go to the “Categories Menu”.
Tap on the category “Protection Options”.
Tap on the category “Setpoint Limits”.
Tap on the sub-category “Minimum Setpoint”.
Enter the new value, using the numeric keypad that appears.
Confirm your entry by tapping on “OK”.
In the display that follows, confirm your entry again by tapping on “OK”.
The correct selection will be displayed graphically and the “Min. setpoint” will be changed
promptly. If tapping on “OK” is not correct, it will be displayed graphically for 2 seconds. The dis-
play will then return to the category “Setpoint Limits”. Try changing the “Min. Setpoint” again.
Tap on the sub-category “Maximum Setpoint”.
Enter the new value, using the numeric keypad that appears.
Confirm your entry by tapping on “OK”.
In the display that follows, confirm your entry again by tapping on “OK”.
The correct selection will be displayed graphically and the “Max. Setpoint” will be changed
promptly. If tapping on “OK” is not correct, it will be displayed graphically for 2 seconds. The dis-
play will then return to the category “Setpoint Limits”. Try changing the “Max. Setpoint” again.
Check the set values for the minimum and maximum setpoint at any system change, especially
when changing the thermal fluid.
4.2.13
Setting the setpoint
PROCEDURE
Go to the “Home” screen.
Tap on the keypad symbol next to “T
Setpoint
”.
Enter a new setpoint using the number keypad that appears.
The following must apply:
[Minimum setpoint limit] ≤ [Setpoint] ≤ [Maximum setpoint limit].
If these conditions are not met, a message will appear on the
>Touchscreen<
[88] and the entry
will be ignored. In this case, delete the value just entered either with the “Arrow” key or with the
“clear” key. Enter the setpoint again.
Confirm your entry by tapping on “OK”.
Overview of the
temperature thresh-
olds
Содержание Pilot ONE Unichiller 00 eo Series
Страница 1: ...Unichiller eo Table Top Models...
Страница 2: ......
Страница 3: ...OPERATION MANUAL Unichiller eo Table Top Models...
Страница 4: ......
Страница 6: ...OPERATION MANUAL Unichiller eo Desktop V2 2 0en 09 08 21 17 12 6 Pilot ONE Layout of the Home screen...
Страница 85: ...Annex Chapter 9 OPERATION MANUAL V2 2 0en 09 08 21 17 12 Unichiller eo Desktop 85 9 Annex...