Commissioning
OPERATION MANUAL
Chapter 2
Unichiller® eo Desktop
V2.2.0en/09.08.21//17.12
30
2.8
Preparations for operation
2.8.1
Unscrewing/activating the leveling feet (if any)
The leveling feet are not unscrewed/activated before switching on the temperature control unit
DEATH OR SERIOUS INJURY DUE TO CRUSHING
The parking brakes must be activated at the casters (if any) and/or the leveling feet must be
unscrewed/activated before the temperature control unit is put into operation.
The temperature control unit may move if the parking brakes of the casters (if any) are not
activated and/or the leveling feet are not unscrewed/activated.
Always unscrew/activate the leveling feet before switching on the temperature control unit.
Uneven floors can be compensated by adjusting these leveling feet.
PROCEDURE
Verify that the parking brakes of the casters (if any) have been activated.
Unscrew the leveling feet.
Compensate uneven floors by adjusting these leveling feet, if necessary. Use a spirit level to
horizontally align the temperature control unit.
Tighten the lock screws on the leveling feet after aligning the temperature control unit. This
prevents the leveling feet from changing their height during operation.
2.8.2
Opening/closing the bypass valve
Some temperature control units are fitted with an adjustable bypass to protect fragile applications
(e.g. a glass apparatus). To find out whether your temperature control unit is equipped with an
adjustable bypass, refer to the “Wiring diagram”.
→
The
>Bypass valve<
[62] is located on top of the temperature control unit. The set pressure is indi-
cated on the
>Touchscreen<
[88] in [Field 8]. The
>Bypass valve<
[62] must be fully open before the
circulation starts:
▪
at the initial filling of the machine;
▪
when switching to another thermal fluid;
▪
when switching to another application.
Opening the bypass valve:
Open the valve by turning it counterclockwise (turn 90° left as f
ar as it will go).
Closing the bypass valve:
Close the valve by turning it clockwise (turn 90° right as far as it will go).
PROCEDURE
Check whether the
>Bypass valve<
[62] is open.
Open the
>Bypass valve<
[62]
by turning it counterclockwise (turn 90° left as far as it will go).
2.8.3
Installing collecting container
PROCEDURE
Install a suitable hose at the
>overflow<
[12] on the temperature control unit (if present). The
hose must be compatible with the thermal fluid and the temperature.
Put the other end of the hose in a suitable collecting container.
Opening and closing
the bypass valve
Содержание Pilot ONE Unichiller 00 eo Series
Страница 1: ...Unichiller eo Table Top Models...
Страница 2: ......
Страница 3: ...OPERATION MANUAL Unichiller eo Table Top Models...
Страница 4: ......
Страница 6: ...OPERATION MANUAL Unichiller eo Desktop V2 2 0en 09 08 21 17 12 6 Pilot ONE Layout of the Home screen...
Страница 85: ...Annex Chapter 9 OPERATION MANUAL V2 2 0en 09 08 21 17 12 Unichiller eo Desktop 85 9 Annex...