34
Press the UP button on the front panel, the time preset value will plus 1; Press the up
button continuously and hold the button, the time preset value will increase 1 continuously.
If a certain value is added to, the time preset value will increase 10 until up to 999s.
②
Decrease
Press the DOWN button on the front panel, the time preset value will minus 1; Press the
down button continuously and hold the button, the time preset value will minus 1
continuously. If a certain value is decreased to, the time preset value will decrease 10 until
up to 0.0s.
7.
Preset the ACW leak current alarm value
(
1
)
Preset conditions
Pop-up the AC / DC switch button in the front panel to make the tester be on the ACW test
status. Note: If the tester is single ACW test, this step can be ignored.
(
2
)
Preset methods
Press the test / preset buttons, the LED window show the current pre-set value. Adjust the
potentiometer clockwise, the current pre-set value will be increased. Counterclockwise
adjustment will reduce the pre-set leak current. If the preset current is up to the desired
value, pop-up the test / preset button.
8.
Preset the DCW leak current alarm value (Single ACW tester is
without DCWtest)
(
1
)
Preset conditions
Press the AC / DC switch button in the front panel to make the tester be on the DCW test
status.
(
2
)
Preset methods
Press the test / preset buttons, the LED window show the current pre-set value. Adjust the
potentiometer clockwise, the current pre-set value will be increased. Counterclockwise
adjustment will reduce the pre-set leak current. If the preset current is up to the desired
value, pop-up the test / preset button.
(
3
)
Adjust the output voltage