33
4AX)
(
1
)
The voltage change-over switch is pressed, and the DC step is set;
(
2
)
The high voltage line of the measured object connects to the high voltage DC output
terminal;
(
3
)
Continue to operate according to the above the required testing value of leakage
current set 3 and 4;
(
4
)
Caution: voltage of the high voltage DC output terminal is negative voltage;
(
5
)
The setting range of the leakage current alarm value is from 0.3 mA to 20 mA. When
the leakage current is above 20 mA, the current automatic protection and alarm are
activated.
5.Remote Control Test: (Note: When a remote control test probe is
applied, the time is controlled manually.)
Example: The timing time is set to 11 s, and the remote control test gun is used to test the
measured object. The test gun is activated till the time is over, and then the test gun will
restart.
(
1
)
The remote control testing accessories connects to the instrument, and the testing
voltage (AC/DC) is selected;
(
2
)
The required value of testing voltage is set;
(
3
)
The required value of leakage current testing is set;
(
4
)
The remote control test gun connects to the measured object;
(
5
)
The switch “Start” on the high voltage gun is pressed to test, and while the test finish,
the switch is released.
6.
Pre-set the test time
(
1
)
Preset conditions
The tester must be in the reset state, the tester can not be in the test status or the alarm
status.
(
2
)
Preset methods
①
Increase