35
When the voltage regulator knob is at the position of 0, there is no high voltage output
even if starts the tester. Press the start button and adjust the output voltage knob
clockwise slowly until the output voltage reaches to the required voltage
VIII.Applying Illustration
1.Preparation of test
①
Avoid electric shock
Please wear insulated gloves and stand on the insulation pad when operate the hipot
tester. Note: the withstanding voltage of insulating gloves and insulated pad should
be twice than the maximum output voltage of the tester.
②
Set the alarm value of leak current and test time
Please refer to the seventh chapter.
③
Connect the DUT(Device Under Test) to the tester
Please confirm the following before connection:
1, The tester is shut down or on the reset status; 2, The high voltage indicator is not light;
3, The voltage window show 0;
Connect the current measurement loop terminal firstly then connect DUT to the high
voltage terminal.
2.Test mode
①
Manual test
If the test time is set to 0, the tester will not judge the test time during testing. The tester is
in a continuous test condition in the test.
After the "START" button is pressed, the test light will be lit; adjust the output voltage knob
at a desired value, the tester will do the test continuously. During the test, if the test
current is great than the preset value, the tester will alarm with sound and light. Press the
"STOP" key, the alarm will be stopped. Replace the DUT and continue to test. After testing,
press the “STOP” key to stop the test.
②
Automatic test
If the test time is not set to 0, the tester will enter into the automatic test. After the "START"
button is pressed, the test light will be lit; adjust the output voltage knob at a desired value,