Adjustments
Power Reference Oscillator Level Adjustment
HP EPM-441A/442A Service Guide
3-9
8.
Observe the reading on the DVM. If less than 400
µ
V, proceed with
the next step. If 400
µ
V or greater, press and hold the HP 432A
fine zero switch and adjust the coarse zero control so that the
DVM indicates 200
µ
V or less. Then release the fine zero switch
and proceed to the next step.
9.
Round off the DVM reading to the nearest microvolt and record
this value as V
0
.
■
V
0
________________
µ
V
■
Disconnect the DVM negative input lead from the V
RF
connector on the test power meter and reconnect it to chassis
ground.
4402serv.book Page 9 Monday, March 11, 2002 11:34 AM
Содержание E4418A
Страница 3: ...Service Guide HP EPM E4418A E4419A EPM 441A 442A Power Meters HP Part no E4418 90006 January 24 1997 ...
Страница 16: ...About this Guide xiv HP EPM 441A 442A Service Guide 4402serv book Page xiv Monday March 11 2002 11 34 AM ...
Страница 20: ...Contents 4 HP EPM 441A 442A Service Guide 4402serv book Page 4 Monday March 11 2002 11 34 AM ...
Страница 22: ...Contents 6 HP EPM 441A 442A Service Guide 4402serv book Page 6 Monday March 11 2002 11 34 AM ...
Страница 24: ...Contents 8 HP EPM 441A 442A Service Guide 4402serv book Page 8 Monday March 11 2002 11 34 AM ...
Страница 25: ...1 Specifications 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Страница 41: ...2 Performance Tests 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Страница 59: ...3 Adjustments 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Страница 71: ...4 Theory of Operation 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Страница 83: ...HP EPM 441A 442A Service Guide Instrument Block Diagram pullout1 fm Page 1 Monday March 11 2002 1 39 PM ...
Страница 85: ...5 Replaceable Parts 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Страница 108: ...Index 4 HP EPM 441A 442A Service Guide 4402serv book Page 4 Monday March 11 2002 11 34 AM ...