Performance Tests
Instrument Accuracy Test
HP EPM-441A/442A Service Guide
2-7
Instrument Accuracy Test
Specification
Description
The power meter accuracy is verified for various power inputs. The range
calibrator is switched to provide these reference inputs.
Equipment
Range Calibrator ................................... HP 11683A
Power Sensor Cable .............................. HP 11730A
Test Setup
Figure 2-2: Instrument Accuracy Test Setup
Electrical Characteristics
Performance Limits
Accuracy
±
0.5% or ±0.02 dB
1
1. This performance limit does not include the corresponding sensor power
linearity specification.
HP 11730A
POWER
METER
CHANNEL A
Power Meter
Range
Calibrator
4402serv.book Page 7 Monday, March 11, 2002 11:34 AM
Содержание E4418A
Страница 3: ...Service Guide HP EPM E4418A E4419A EPM 441A 442A Power Meters HP Part no E4418 90006 January 24 1997 ...
Страница 16: ...About this Guide xiv HP EPM 441A 442A Service Guide 4402serv book Page xiv Monday March 11 2002 11 34 AM ...
Страница 20: ...Contents 4 HP EPM 441A 442A Service Guide 4402serv book Page 4 Monday March 11 2002 11 34 AM ...
Страница 22: ...Contents 6 HP EPM 441A 442A Service Guide 4402serv book Page 6 Monday March 11 2002 11 34 AM ...
Страница 24: ...Contents 8 HP EPM 441A 442A Service Guide 4402serv book Page 8 Monday March 11 2002 11 34 AM ...
Страница 25: ...1 Specifications 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Страница 41: ...2 Performance Tests 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Страница 59: ...3 Adjustments 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Страница 71: ...4 Theory of Operation 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Страница 83: ...HP EPM 441A 442A Service Guide Instrument Block Diagram pullout1 fm Page 1 Monday March 11 2002 1 39 PM ...
Страница 85: ...5 Replaceable Parts 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Страница 108: ...Index 4 HP EPM 441A 442A Service Guide 4402serv book Page 4 Monday March 11 2002 11 34 AM ...