Adjustments
Power Reference Oscillator Level Adjustment
HP EPM-441A/442A Service Guide
3-11
Measurement Uncertainty
When calculating measurement uncertainty the significant multimeter
uncertainty parameters are for V
comp
, V
1
-V
0
and R. Other uncertainties
which should be included are:
•
Maths Assumptions:
±
0.01%
•
Effective Efficiency Calibration:
±
0.05%
•
Mismatch Uncertainty with Source and Mount:
±
0.1%
•
SWR:
≤1.05
These values are typical.
4402serv.book Page 11 Monday, March 11, 2002 11:34 AM
Содержание E4418A
Страница 3: ...Service Guide HP EPM E4418A E4419A EPM 441A 442A Power Meters HP Part no E4418 90006 January 24 1997 ...
Страница 16: ...About this Guide xiv HP EPM 441A 442A Service Guide 4402serv book Page xiv Monday March 11 2002 11 34 AM ...
Страница 20: ...Contents 4 HP EPM 441A 442A Service Guide 4402serv book Page 4 Monday March 11 2002 11 34 AM ...
Страница 22: ...Contents 6 HP EPM 441A 442A Service Guide 4402serv book Page 6 Monday March 11 2002 11 34 AM ...
Страница 24: ...Contents 8 HP EPM 441A 442A Service Guide 4402serv book Page 8 Monday March 11 2002 11 34 AM ...
Страница 25: ...1 Specifications 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Страница 41: ...2 Performance Tests 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Страница 59: ...3 Adjustments 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Страница 71: ...4 Theory of Operation 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Страница 83: ...HP EPM 441A 442A Service Guide Instrument Block Diagram pullout1 fm Page 1 Monday March 11 2002 1 39 PM ...
Страница 85: ...5 Replaceable Parts 4402serv book Page 1 Monday March 11 2002 11 34 AM ...
Страница 108: ...Index 4 HP EPM 441A 442A Service Guide 4402serv book Page 4 Monday March 11 2002 11 34 AM ...