Monitoring
Using Leads to Monitor
3-3
5-Wire Patient Cable
Table 3-4, “5-Wire Electrode Placement,” on page 3-3 describes a typical lead-wire
placement using the 5-wire patient cable. Table 3-5, “Lead Configurations,” on page
3-3 shows how the individual leads are formed using the individual leadwires.
Table 3-3
Lead Formation
Lead
+
—
ref
I
LA
RA
LL
II
LL
RA
LA
III
LL
LA
RA
Table 3-4
5-Wire Electrode Placement
Electrode
Placement
RA/White
Near right midclavicular line, directly below clavicle.
LA/Black
Near left midclavicular line, directly below clavicle.
LL/Red
Below the left pectoral muscle on the left midclavicular line.
RL/Green
Below the right pectoral muscle on the right midclavicular line.
V/Brown
As appropriate for the V lead to be monitored (V1 - V6).
Table 3-5
Lead Configurations
Lead
Leadwire Combinations
I
LA - RA
II
LL - RA
III
LL - LA
aVR
RA - .5 (LA + LL)
aVF
LL - .5 (RA + LA)