Troubleshooting
Troubleshooting
7-5
Troubleshooting the Pacer
Table 7-3
Pacer Messages
Message
Cause
Possible Solutions
PACER
FAILURE
The unit detected a delivered current
error.
1
Change pads and make
sure pads adaptor cable is
properly connected.
2
Turn the Energy Select
switch to Off (Standby),
then turn back to Monitor
On.
3
Restart the pacer.
If
PACER FAILURE
hap-
pens again, use a back-up
pacer and call for service.
PACER OUT-
PUT LOW
Pacer current
highlighted on
the display.
The pacer cannot deliver the
required current.
1
Stop the pacer.
2
Change the pads and
check the pads adaptor
cable for proper connec-
tion.
3
Restart the pacer.
STOP PACER
An attempt was made to change pac-
ing mode while pacing.
Stop the pacer before you
change the pacing mode.
NO PADS
Pacer is on and one of the following
occurs:
•
No pads adaptor attached.
•
Paddles attached.
•
Internal paddles attached.
Attach pads for pacing.
PADS OFF
Pads adaptor cable is attached and a
pad is off.
Attach pad for pacing. If pads
are already applied to patient,
change to a new set of pads.
ATTACH
PADS
Pads adaptor cable is attached.
is pressed and pads are
not attached.
is pressed
and pads are not attached.
Attach the pacer pads.
Charg
Pacer
Содержание CodeMaster XL+ M1722A
Страница 6: ...vi ...
Страница 11: ...List of Tables Contents 6 ...
Страница 34: ...Defibrillating Defibrillating a Patient 2 10 ...
Страница 46: ...Monitoring External Monitoring 3 12 ...
Страница 57: ...Pacing Optional Using the Pacer 5 5 After the shock the pacer remains off Resume pacing if it is required ...
Страница 58: ...Pacing Optional Using the Pacer 5 6 ...
Страница 106: ...SpO2 Monitoring Optional SpO2 Alarms 6 10 ...