6-1
6
SpO
2
Monitoring (Optional)
SpO
2
monitoring gives information on both cardiac and respiratory systems, and
details of oxygen transportation in the body. It is widely used because it is non-
invasive, continuous, easily applied and painless.
The quality of SpO
2
measurements depends on careful application of the sensor.
Read the following section, “Application Notes” (page 6-1), to understand the
importance of sensor application. For more detailed information, refer to SpO
2
Concepts (HP part number M1722-93950).
You can use the SpO
2
monitor with sensors made by other manufacturers as well as
with HP sensors. For a list of approved sensors, see the Sensor Guide (HP part
number M1722-93970).
SaO
2
and SpO
2
Hewlett-Packard is adopting the convention of referring to the SpO
2
parameter.
Previously it was referred to as SaO
2
.
SaO
2
is the term used to indicate the oxygen saturation of arterial blood.
SpO
2
is the term used to indicate the oxygen saturation of arterial blood as
measured by pulse oximetry.
6
Application Notes
The pulse oximetry method used for measuring SpO
2
uses LEDs (light emitting
diodes) to transmit red and infrared light through suitable peripheral areas of the
body, typically the foot in neonates or the finger in adults. The oxygen saturation is
gauged by measuring the "redness" of the blood in the arterial pulse.
A photodetector positioned opposite the light emitter compares light absorption
before and after pulsation to provide measurements that are displayed on the
monitor. If there is no pulse, measurements cannot be made. See the following
figure.
Содержание CodeMaster XL+ M1722A
Страница 6: ...vi ...
Страница 11: ...List of Tables Contents 6 ...
Страница 34: ...Defibrillating Defibrillating a Patient 2 10 ...
Страница 46: ...Monitoring External Monitoring 3 12 ...
Страница 57: ...Pacing Optional Using the Pacer 5 5 After the shock the pacer remains off Resume pacing if it is required ...
Страница 58: ...Pacing Optional Using the Pacer 5 6 ...
Страница 106: ...SpO2 Monitoring Optional SpO2 Alarms 6 10 ...