SpO2 Monitoring (Optional)
SpO2 Alarms
6-8
The pulse rate is derived from the pulse oximeter. It should correlate closely with
the patient’s heart rate.
6
SpO
2
Alarms
Activating SpO
2
Alarms
There are three preset high/low SpO
2
alarm limits: 100/90, 100/85 and 100/80. Press
the
button repeatedly to cycle through the alarm options and the no-
alarm option. Stop when you see the alarm you would like to choose and after three
seconds that alarm will take effect. A
symbol replaces the limits to show that
the alarm is active. To review the limit set, press the
button.
If the SpO
2
level falls below the low alarm limit, an alarm sounds and the violated
limit is highlighted.
SpO
2
alarms are automatically turned off when you press
.
Deactivating SpO
2
Alarms
Press the
button. The
symbol to the right of the SpO
2
display
indicates that the alarms are off.
SpO
2
Alarm
SpO
2
Alarm
NOTE
Charge
SpO
2
Alarm