Defibrillating
Defibrillating a Patient
2-3
Wait for the Charge Done indicators. When the unit is armed, the monitor Delivered
Energy display shows the available energy in joules.
If the defibrillator does not charge, refer to Chapter 7, Troubleshooting.
Resetting the Selected Energy Level
To increase or decrease the selected energy level after pressing the Charge button,
perform the following steps.
1
Move the Energy Select control to the new energy level.
2
Wait for the Charge Done indicators.
3. Shock
To shock the patient, perform the following steps.
1
Briefly adjust paddle pressure and placement to optimize patient contact, as reg-
istered on the paddle contact indicator (if supplied).
2
Verify that no one is in contact with the patient, the monitoring cable or leads,
the bed rails, or another potential current pathway.
3
Call out "Clear!" to alert personnel to stand away from the patient.
Keep hands clear of the paddle electrode edges. Use your thumbs
to depress the Shock buttons on the paddle handles.
4
Press and briefly hold both Shock buttons (one on each paddle) simultaneously,
to deliver energy to the patient.
If the defibrillator does not shock, refer to Chapter 7, Troubleshooting.
If you must disarm the charged defibrillator (if countershock is not needed), turn the
Energy Select control to Monitor On. Any stored energy will be discharged
internally and the available energy on the display will return to 0.
WARNING
NOTE
Содержание CodeMaster XL+ M1722A
Страница 6: ...vi ...
Страница 11: ...List of Tables Contents 6 ...
Страница 34: ...Defibrillating Defibrillating a Patient 2 10 ...
Страница 46: ...Monitoring External Monitoring 3 12 ...
Страница 57: ...Pacing Optional Using the Pacer 5 5 After the shock the pacer remains off Resume pacing if it is required ...
Страница 58: ...Pacing Optional Using the Pacer 5 6 ...
Страница 106: ...SpO2 Monitoring Optional SpO2 Alarms 6 10 ...