Defibrillating a Patient
3-7
Defibrillating through Multifunction Defib Electrodes
The HP CodeMaster 100 has a multifunction electrode adapter cable (M2472A).
This adapter cable allows defibrillation through multifunction defib electrodes.
Defib electrodes have several advantages:
• Allows "hands off" defibrillation.
• Provides good quality monitoring.
• You can perform synchronized cardioversion without using an ECG lead set, while
monitoring through the electrodes.
• You can switch between the pacing and defibrillation modes of operation quickly.
The defibrillator will deliver defibrillator energy levels to an open
electrodes set. The message
PADS OFF
appears when there is poor
electrode-to-patient contact. Check all patient connections if this message
appears.
1
Attach the electrode adapter cable (M2472A) to the paddle connector (Figure 3-
2) on the front of the defibrillator.
2
Slide the cable connector latch to the lock position to secure the cable.
3
Attach the multifunction defib electrodes to the patient as instructed on the
package.
4
Connect the defib electrodes to the electrodes adapter cable. The defib elec-
trodes are correctly connected when the locking ring is twisted, locking the ears
of the connector to the adapter cable.
Be sure to correctly connect the defib electrodes to the electrodes adapter
cable. Incorrectly connecting the defib electrodes to the electrode adapter
cable can result in a failure to deliver energy to the patient.
WA RNI N G
WA RNI N G
Содержание Codemaster 100
Страница 18: ...Telecom Regulatory Information xviii ...
Страница 28: ...xxviii ...
Страница 54: ... HWWLQJ FTXDLQWHG Battery Operation 1 26 ...
Страница 84: ...6HWXS DQG RQILJXUDWLRQ Configuring the 12 Lead Option 2 30 ...
Страница 94: ... HILEULOODWLQJ Defibrillating a Patient 3 10 ...
Страница 133: ...8VLQJ WKH HDG 2SWLRQ 12 Lead Keyboard and Display 6 5 Figure 6 2 Characters by Language 6 2 ...
Страница 150: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 22 Figure 6 4 The ECG Report 6 4 0 4 3 6 5 2 7 8 ...
Страница 154: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 26 Figure 6 5 A Standard 3x4 ECG 3x4 6 5 ...
Страница 155: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 27 Figure 6 6 3x4 ECG with One Rhythm Strip 3x4 1R 6 6 ...
Страница 156: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 28 Figure 6 7 An Auto 3x4 ECG with Three Rhythm Strips 3x4 3R 6 7 ...
Страница 178: ...3DFLQJ Using the Pacer 8 8 ...
Страница 202: ...7UDQVPLWWLQJ DWD IURP WKH HDG 2SWLRQ Transmitting an Event Summary 10 14 ...
Страница 234: ...7URXEOHVKRRWLQJ Service 12 22 ...