Configuring the 12-Lead Option
2-23
Setting Copy Count
Determines how many copies are printed when you press the Copy key. You can
enter a value from 1-4. The default value is 2.
Setting Keyboard Values
Keyboard mode determines whether characters will be entered in uppercase or
lowercase. Press F3 to change from lowercase to UPPERCASE and vice versa. The
keyboard is initially set to lowercase. Use the shift key to enter the opposite case.
Setting 12-Lead Location Code
The 12-Lead location code identifies a specific department or EMS Service that
acquired the ECG. The number prints at the bottom of the report. If set to zero
(default), this field will be suppressed from the printed report. This parameter can be
any five digit numeric code.
Setting 12-Lead ID
The 12-Lead ID code identifies the particular CodeMaster 100 unit that was used to
acquire the ECG. The number prints at the bottom of the report. If set to zero
(default), this field will be omitted from the printed report. This parameter can be
any four digit numeric code.
Содержание Codemaster 100
Страница 18: ...Telecom Regulatory Information xviii ...
Страница 28: ...xxviii ...
Страница 54: ... HWWLQJ FTXDLQWHG Battery Operation 1 26 ...
Страница 84: ...6HWXS DQG RQILJXUDWLRQ Configuring the 12 Lead Option 2 30 ...
Страница 94: ... HILEULOODWLQJ Defibrillating a Patient 3 10 ...
Страница 133: ...8VLQJ WKH HDG 2SWLRQ 12 Lead Keyboard and Display 6 5 Figure 6 2 Characters by Language 6 2 ...
Страница 150: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 22 Figure 6 4 The ECG Report 6 4 0 4 3 6 5 2 7 8 ...
Страница 154: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 26 Figure 6 5 A Standard 3x4 ECG 3x4 6 5 ...
Страница 155: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 27 Figure 6 6 3x4 ECG with One Rhythm Strip 3x4 1R 6 6 ...
Страница 156: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 28 Figure 6 7 An Auto 3x4 ECG with Three Rhythm Strips 3x4 3R 6 7 ...
Страница 178: ...3DFLQJ Using the Pacer 8 8 ...
Страница 202: ...7UDQVPLWWLQJ DWD IURP WKH HDG 2SWLRQ Transmitting an Event Summary 10 14 ...
Страница 234: ...7URXEOHVKRRWLQJ Service 12 22 ...