6HWXSDQG&RQILJXUDWLRQ
Configuring the 12-Lead Option
2-19
Setting Up Filters
From the Configuration Menu, select Setup ECG Filters. The 12-Lead option has
been factory-configured with the filter settings which remove the most noise from
the ECG. In addition to the default settings, 12-Lead offers a choice of several filter
configurations. These filter settings are detailed below and in Table 2-5.
• The 0.5 Hz Baseline Wander filter suppresses the greatest amount of baseline
wander.
• The 0.15 Hz Baseline Wander filter provides some baseline wander suppression
without distorting the ECG’s ST segment.
• The 0.05 Hz Baseline Wander filter delivers the highest fidelity signal, but pro-
vides the least baseline wander suppression.
• The 40 Hz Noise filter offers maximum noise suppression, but reduces the fidelity
of the signal.
• The 100 Hz Noise filter provides some noise suppression while offering an accu-
rate signal representation.
• The 150 Hz Noise filter delivers the highest fidelity signal, but provides the least
high-frequency noise suppression.
Setup ECG Filters
12-Lead Report Filters -
Wander Filter:
1.5 Hz
Noise Filter:
40 Hz
Front Panel Filter Key-
Muscle Artifact Filter:
Yes
40 Hz Noise Filter:
Yes
0.5 Hz Wander Filter:
Yes
Change Value
Exit
F1
F2
F3
F4
F5
Содержание Codemaster 100
Страница 18: ...Telecom Regulatory Information xviii ...
Страница 28: ...xxviii ...
Страница 54: ... HWWLQJ FTXDLQWHG Battery Operation 1 26 ...
Страница 84: ...6HWXS DQG RQILJXUDWLRQ Configuring the 12 Lead Option 2 30 ...
Страница 94: ... HILEULOODWLQJ Defibrillating a Patient 3 10 ...
Страница 133: ...8VLQJ WKH HDG 2SWLRQ 12 Lead Keyboard and Display 6 5 Figure 6 2 Characters by Language 6 2 ...
Страница 150: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 22 Figure 6 4 The ECG Report 6 4 0 4 3 6 5 2 7 8 ...
Страница 154: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 26 Figure 6 5 A Standard 3x4 ECG 3x4 6 5 ...
Страница 155: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 27 Figure 6 6 3x4 ECG with One Rhythm Strip 3x4 1R 6 6 ...
Страница 156: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 28 Figure 6 7 An Auto 3x4 ECG with Three Rhythm Strips 3x4 3R 6 7 ...
Страница 178: ...3DFLQJ Using the Pacer 8 8 ...
Страница 202: ...7UDQVPLWWLQJ DWD IURP WKH HDG 2SWLRQ Transmitting an Event Summary 10 14 ...
Страница 234: ...7URXEOHVKRRWLQJ Service 12 22 ...