7URXEOHVKRRWLQJ
CodeMaster 100
12-4
If the Defibrillator does not Deliver a Shock
If the defibrillator does not deliver a shock, perform the following steps:
1
Make sure the unit is not in synchronized shock mode by checking that
the
Sync
light is not on.
2
If the unit discharges internally (that is, the energy display decrements
slowly then beeps three times and displays the screen message
DEFIB
DISARMED
) verify proper connections from the patient to the defibrilla-
tor. This includes connections to the pads/paddles adapter cable if one is
being used. Also check for worn or broken areas along the cables.
3
Press
Charge
again, wait for the Charge Done indicators.
4
Press the
Shock
(
) buttons again.
5
If the unit remains unable to shock, turn the
Energy Select
control to
the
Monitor On
or
Off
position and use a back-up defibrillator.
6
Press
Review
to print an event summary and keep any ECG strips from
the defibrillator for later evaluation.
7
Contact HP service personnel as soon as possible.
Содержание Codemaster 100
Страница 18: ...Telecom Regulatory Information xviii ...
Страница 28: ...xxviii ...
Страница 54: ... HWWLQJ FTXDLQWHG Battery Operation 1 26 ...
Страница 84: ...6HWXS DQG RQILJXUDWLRQ Configuring the 12 Lead Option 2 30 ...
Страница 94: ... HILEULOODWLQJ Defibrillating a Patient 3 10 ...
Страница 133: ...8VLQJ WKH HDG 2SWLRQ 12 Lead Keyboard and Display 6 5 Figure 6 2 Characters by Language 6 2 ...
Страница 150: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 22 Figure 6 4 The ECG Report 6 4 0 4 3 6 5 2 7 8 ...
Страница 154: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 26 Figure 6 5 A Standard 3x4 ECG 3x4 6 5 ...
Страница 155: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 27 Figure 6 6 3x4 ECG with One Rhythm Strip 3x4 1R 6 6 ...
Страница 156: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 28 Figure 6 7 An Auto 3x4 ECG with Three Rhythm Strips 3x4 3R 6 7 ...
Страница 178: ...3DFLQJ Using the Pacer 8 8 ...
Страница 202: ...7UDQVPLWWLQJ DWD IURP WKH HDG 2SWLRQ Transmitting an Event Summary 10 14 ...
Страница 234: ...7URXEOHVKRRWLQJ Service 12 22 ...