3DFLQJ
Using the Pacer
8-4
Demand Mode Pacing When in demand mode, the pacer will only deliver pacer
pulses when the patient’s heart rate is lower than the selected pacer rate.
Fixed Mode Pacing When in fixed mode, the pacer will deliver pacer pulses at the
selected pacer rate.
Use demand pacing mode whenever possible. Use fixed pacing mode
(asynchronous) for cases when reliably monitoring the patient is
impractical. For example, use fixed mode when there is motion artifact or
other ECG noise that makes R-wave detection unreliable.
12 Press Start/Stop to start pacing. The monitor displays the message
PACING
as
well as the selected mode, rate, and output.
The pacer will not start pacing if there is a problem with either the multifunction
defib electrode connections or the ECG monitoring electrode connections. If
there is a problem with the multifunction defib electrode connection, the mes-
sage
ATTACH PADS
will be displayed briefly when you press Start/Stop.
13 Verify that the pacer pulses are well positioned in the diastole.
14 Increase output (mA) by pressing Output until the beat is captured. Selecting an
alternate lead can help you determine capture.
WA RNI N G
Содержание Codemaster 100
Страница 18: ...Telecom Regulatory Information xviii ...
Страница 28: ...xxviii ...
Страница 54: ... HWWLQJ FTXDLQWHG Battery Operation 1 26 ...
Страница 84: ...6HWXS DQG RQILJXUDWLRQ Configuring the 12 Lead Option 2 30 ...
Страница 94: ... HILEULOODWLQJ Defibrillating a Patient 3 10 ...
Страница 133: ...8VLQJ WKH HDG 2SWLRQ 12 Lead Keyboard and Display 6 5 Figure 6 2 Characters by Language 6 2 ...
Страница 150: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 22 Figure 6 4 The ECG Report 6 4 0 4 3 6 5 2 7 8 ...
Страница 154: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 26 Figure 6 5 A Standard 3x4 ECG 3x4 6 5 ...
Страница 155: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 27 Figure 6 6 3x4 ECG with One Rhythm Strip 3x4 1R 6 6 ...
Страница 156: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 28 Figure 6 7 An Auto 3x4 ECG with Three Rhythm Strips 3x4 3R 6 7 ...
Страница 178: ...3DFLQJ Using the Pacer 8 8 ...
Страница 202: ...7UDQVPLWWLQJ DWD IURP WKH HDG 2SWLRQ Transmitting an Event Summary 10 14 ...
Страница 234: ...7URXEOHVKRRWLQJ Service 12 22 ...