CodeMaster 100
13-13
A new HP M2477B 4.0 amp-hour battery will provide 75 360-joule
charge-shock cycles or a minimum of 3 hours of monitoring time. Replace the
battery when it fails to provide the minimum3 hours of monitoring time or 10
minutes of Low Battery warning time (starting from a fully charged battery).
Properly dispose of depleted batteries according to local regulations. Do
not disassemble, puncture or incinerate the disposed batteries. Be careful
not to short the battery terminals, because this can cause a fire hazard.
New Batteries
To assure optimal performance of a new battery, all newly purchased
batteries should be capacity tested in the M2480B charger prior to being put
into service. Accomplish this by placing the battery in the charger and
performing the “Battery Capacity Test” on page 13-15.
Battery Life
Recommended battery pack replacement:
• After two years of service.
• After failure of battery capacity test.
Replace the battery only with one of the battery packs listed in Table 1-2 on
page 1-24.
The operating time of the battery depends on how much the battery pack is
charged. Press the capacity gauge button on the battery pack to determine
the charge level.
NO TE
WA RNI N G
Содержание Codemaster 100
Страница 18: ...Telecom Regulatory Information xviii ...
Страница 28: ...xxviii ...
Страница 54: ... HWWLQJ FTXDLQWHG Battery Operation 1 26 ...
Страница 84: ...6HWXS DQG RQILJXUDWLRQ Configuring the 12 Lead Option 2 30 ...
Страница 94: ... HILEULOODWLQJ Defibrillating a Patient 3 10 ...
Страница 133: ...8VLQJ WKH HDG 2SWLRQ 12 Lead Keyboard and Display 6 5 Figure 6 2 Characters by Language 6 2 ...
Страница 150: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 22 Figure 6 4 The ECG Report 6 4 0 4 3 6 5 2 7 8 ...
Страница 154: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 26 Figure 6 5 A Standard 3x4 ECG 3x4 6 5 ...
Страница 155: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 27 Figure 6 6 3x4 ECG with One Rhythm Strip 3x4 1R 6 6 ...
Страница 156: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 28 Figure 6 7 An Auto 3x4 ECG with Three Rhythm Strips 3x4 3R 6 7 ...
Страница 178: ...3DFLQJ Using the Pacer 8 8 ...
Страница 202: ...7UDQVPLWWLQJ DWD IURP WKH HDG 2SWLRQ Transmitting an Event Summary 10 14 ...
Страница 234: ...7URXEOHVKRRWLQJ Service 12 22 ...