T
est
14.
Miscellaneous
Residual
Responses
The
purpose
of
this
verication
test
is
to
measure
residual
responses
generated
by
the
DUT
with
no
signal
applied
to
the
DUT
RF
INPUT
.
(The
residual
responses
measured
are
harmonics
of
the
300
MHz
reference
input.)
This
is
a
nal
test.
1.
Connect
the
equipment
as
shown
in
the
test
setup
illustration
for
this
procedure
.
2.
Set
the
mainframe
line
switch
to
ON.
3.
Load
and
run
T
est
14.
Miscellaneous
Residual
Responses
.
Refer
to
Chapter
2
for
information
about
running
the
software
.
When
performing
this
verication
test,
before
measuring
residual
responses
,
the
test
characterizes
DUT
gain,
DUT
input
attenuator
,
and
the
calibrated
spectrum
analyzer
input
attenuator
.
The
characterization
data
later
corrects
the
measured
residual
responses
.
Characterizing
the
DUT
Gain
The
full
microwave
source
provides
a
300
MHz,
010
dBm ,
signal
at
the
DUT
RF
INPUT
.
DUT
input
attenuation
is
10
dB.
The
local
oscillator
source
tunes
to
center
the
response
.
The
calibrated
spectrum
analyzer
stores
the
21.4
MHz
OUT
amplitude
and
corrects
it
for
DUT
atness
.
The
test
then
calculates
DUT
gain,
the
corrected
21.4
MHz
amplitude
minus
10
dBm
RF
INPUT
.
Characterizing
the
DUT
Input
Attenuator
With
the
DUT
input
attenuator
set
to
10
dB
attenuation,
the
calibrated
spectrum
analyzer
measures
the
21.4
MHz
OUT
amplitude
.
The
measurement
repeats
with
0
dB
input
attenuation.
The
DUT
attenuator
correction
factor
equals
the
second
measurement
minus
the
rst.
Only
the
0
and
10
dB
DUT
input
attenuator
settings
are
of
interest.
Characterizing
the
Calibrated
Spectrum
Analyzer
Input
Attenuator
With
the
calibrated
spectrum
analyzer
input
attenuator
set
to
10
dB,
the
calibrated
spectrum
analyzer
measures
21.4
MHz
OUT
amplitude
of
the
DUT
.
The
measurement
repeats
with
0
dB
attenuation.
The
calibrated
spectrum
analyzer
attenuation
correction
factor
equals
the
second
measurement
minus
the
rst.
Measuring
Residuals
The
full
microwave
source
RF
output
is
o.
The
calibrated
spectrum
analyzer
settings
are
050
dBm
reference
level,
100
Hz
span,
30
Hz
resolution
bandwidth.
The
DUT
input
attenuation
is
0
dB.
In
1H0
band
only
,
the
local
oscillator
source
tunes
in
sequence
to
the
fundamental
and
the
rst
eight
harmonics
of
300
MHz,
and
the
calibrated
spectrum
analyzer
measures
responses
at
the
21.4
MHz
OUT
.
The
test
corrects
the
measurements
for
DUT
gain,
DUT
attenuation
error
,
calibrated
spectrum
analyzer
error
,
and
DUT
atness
,
and
then
compares
the
results
with
test
limits
.
8-30
Module
V
erication
T
ests
Содержание 70908A
Страница 7: ...Servicing at a Glance vi ...
Страница 19: ......
Страница 44: ...Figure 2 1 Main Menu Softkeys Module Veri cation Software 2 19 ...
Страница 45: ...Figure 2 2 Mass Storage Menu and Parameter Menu Softkeys 2 20 Module Veri cation Software ...
Страница 46: ...Figure 2 3 Equipment Menu and HP MSIB Map Screen Menu Softkeys Module Veri cation Software 2 21 ...
Страница 47: ...Figure 2 4 Test Menu Softkeys 2 22 Module Veri cation Software ...
Страница 55: ......
Страница 61: ......
Страница 88: ...Troubleshooting the A1 Leveling Ampli er Figure 4 2 Notch Filter Waveform Troubleshooting 4 27 ...
Страница 94: ...Troubleshooting the A3 SYTFMD Figure 4 5 Discriminator Output Waveform Troubleshooting 4 33 ...
Страница 117: ...Overall Block Diagram of RF Section Back page for Foldout Goes Here This is just a place holder 4 56 Troubleshooting ...
Страница 125: ......
Страница 128: ...External Frequency Reference Figure 6 1 Preferred Frequency Reference Connections System Calibration 6 3 ...
Страница 130: ...External Frequency Reference Figure 6 4 Using an HP 8566B Spectrum Analyzer and a House Standard System Calibration 6 5 ...
Страница 139: ......
Страница 205: ......
Страница 209: ...Test 01 Flatness Figure 8 2 Equipment Setup for Test 01 Flatness 8 4 Module Veri cation Tests ...
Страница 253: ...Test 23 RF Input Return Loss Figure 8 25 Equipment Setup for Test 23 RF Input Return Loss 8 48 Module Veri cation Tests ...
Страница 257: ...Test 25 LO Input Return Loss Figure 8 27 Equipment Setup for Test 25 LO Input Return Loss 8 52 Module Veri cation Tests ...
Страница 267: ......
Страница 271: ...Front Panel Figure 9 2 Front Panel Removal Replacement 1 of 2 9 4 Replacing Major Assemblies ...
Страница 272: ...Front Panel Figure 9 3 Front Panel Removal Replacement 2 of 2 Replacing Major Assemblies 9 5 ...
Страница 274: ...Rear Panel Figure 9 4 Rear Panel Removal Replacement 1 or 2 Replacing Major Assemblies 9 7 ...
Страница 275: ...Rear Panel Figure 9 5 Rear Panel Removal Replacement 2 or 2 9 8 Replacing Major Assemblies ...
Страница 277: ...Bandpass Filter Figure 9 6 Bandpass Filter Removal Replacement 1 or 2 9 10 Replacing Major Assemblies ...
Страница 278: ...Bandpass Filter Figure 9 7 Bandpass Filter Removal Replacement 2 or 2 Replacing Major Assemblies 9 11 ...
Страница 280: ...A1 Leveling Ampli er Figure 9 8 A1 Leveling Ampli er Removal Replacement Replacing Major Assemblies 9 13 ...
Страница 282: ...A2 LO Multiplier Figure 9 9 A2 LO Multiplier Removal Replacement Replacing Major Assemblies 9 15 ...
Страница 284: ...A3 SYTFMD Figure 9 10 A3 SYTFMD Removal Replacement Replacing Major Assemblies 9 17 ...
Страница 286: ...A4 Input Attenuator Figure 9 11 A4 Input Attenuator Removal Replacement 1 of 2 Replacing Major Assemblies 9 19 ...
Страница 287: ...A4 Input Attenuator Figure 9 12 A4 Input Attenuator Removal Replacement 2 of 2 9 20 Replacing Major Assemblies ...
Страница 289: ...A5 VCO Sampler Figure 9 13 A5 VCO Sampler Removal Replacement 1 of 3 9 22 Replacing Major Assemblies ...
Страница 290: ...A5 VCO Sampler Figure 9 14 A5 VCO Sampler Removal Replacement 2 of 3 Replacing Major Assemblies 9 23 ...
Страница 291: ...A5 VCO Sampler Figure 9 15 A5 VCO Sampler Removal Replacement 3 of 3 9 24 Replacing Major Assemblies ...
Страница 294: ...A7 2nd Mixer Figure 9 17 A7 2nd Mixer Removal Replacement Replacing Major Assemblies 9 27 ...
Страница 298: ...A9 2nd Converter PLL Figure 9 19 A9 2nd Converter PLL Removal Replacement 1 of 2 Replacing Major Assemblies 9 31 ...
Страница 299: ...A9 2nd Converter PLL Figure 9 20 A9 2nd Converter PLL Removal Replacement 2 of 2 9 32 Replacing Major Assemblies ...
Страница 301: ...A10 Power Supply Driver Figure 9 21 A10 Power Supply Driver Removal Replacement 9 34 Replacing Major Assemblies ...
Страница 303: ...A11 Last Converter Figure 9 22 A11 Last Converter Removal Replacement 1 of 2 9 36 Replacing Major Assemblies ...
Страница 304: ...A11 Last Converter Figure 9 23 A11 Last Converter Removal Replacement 2 of 2 Replacing Major Assemblies 9 37 ...
Страница 309: ...A15 RF 1st Converter Figure 9 27 A15 RF 1st Converter Removal Replacement 9 42 Replacing Major Assemblies ...
Страница 311: ...A16 Low Pass Filter Figure 9 28 A16 Low Pass Filter Removal Replacement 9 44 Replacing Major Assemblies ...
Страница 313: ...A17 Low Pass Filter Figure 9 29 A17 Low Pass Filter Removal Replacement 9 46 Replacing Major Assemblies ...
Страница 329: ......