A
djustment
24.
Last
Converter
Bandpass
Filter
The
purpose
of
this
adjustment
procedure
is
to
verify
or
adjust
the
321.4
MHz
bandpass
lter
response
of
the
A11
last
converter .
1.
Set
the
mainframe
line
switch
to
OFF
.
2.
Remove
the
HP
70908A
RF
section
from
the
mainframe
.
3.
Install
the
extender
module
in
the
mainframe
and
connect
the
HP
70908A
RF
section
to
the
extender
cable
.
4.
Connect
the
equipment
as
shown
in
Figure
7-50.
5.
Set
the
mainframe
line
switch
to
ON
and
allow
at
least
30
minutes
initial
warmup
before
starting
the
adjustment.
6.
Load
and
run
A
djustment
24.
Last
Converter
Bandpass
Filter
.
Refer
to
Chapter
2
for
information
about
running
the
software
.
A
pulse-modulated
carrier
passes
through
the
directional
bridge
to
the
A11
last-converter
321.4
MHz
HIGH
IF
INPUT
.
The
bridge
and
the
A11
last
converter
output
provide
the
inputs
to
channels
B
and
A
of
the
network
analyzer
.
The
resulting
analyzer
display
is
the
A11
bandpass
lter
response
.
A
prompt
instructs
the
user
to
adjust
C15,
C16,
C17,
C18,
and
C19
for
maximum
bandpass
lter
response
at
278.6
MHz.
(This
initial
adjustment
maximizes
the
following
notch
lter
adjustment.)
The
user
next
adjusts
C19
at
278.6
MHz
for
minimum
power
.
C15,
C16,
C17,
and
C18
are
adjusted
for
best
overall
bandpass
lter
response
at
321.4
MHz.
The
test
veries
the
resulting
bandpass
.
Adjustment
Procedures
7-59
Содержание 70908A
Страница 7: ...Servicing at a Glance vi ...
Страница 19: ......
Страница 44: ...Figure 2 1 Main Menu Softkeys Module Veri cation Software 2 19 ...
Страница 45: ...Figure 2 2 Mass Storage Menu and Parameter Menu Softkeys 2 20 Module Veri cation Software ...
Страница 46: ...Figure 2 3 Equipment Menu and HP MSIB Map Screen Menu Softkeys Module Veri cation Software 2 21 ...
Страница 47: ...Figure 2 4 Test Menu Softkeys 2 22 Module Veri cation Software ...
Страница 55: ......
Страница 61: ......
Страница 88: ...Troubleshooting the A1 Leveling Ampli er Figure 4 2 Notch Filter Waveform Troubleshooting 4 27 ...
Страница 94: ...Troubleshooting the A3 SYTFMD Figure 4 5 Discriminator Output Waveform Troubleshooting 4 33 ...
Страница 117: ...Overall Block Diagram of RF Section Back page for Foldout Goes Here This is just a place holder 4 56 Troubleshooting ...
Страница 125: ......
Страница 128: ...External Frequency Reference Figure 6 1 Preferred Frequency Reference Connections System Calibration 6 3 ...
Страница 130: ...External Frequency Reference Figure 6 4 Using an HP 8566B Spectrum Analyzer and a House Standard System Calibration 6 5 ...
Страница 139: ......
Страница 205: ......
Страница 209: ...Test 01 Flatness Figure 8 2 Equipment Setup for Test 01 Flatness 8 4 Module Veri cation Tests ...
Страница 253: ...Test 23 RF Input Return Loss Figure 8 25 Equipment Setup for Test 23 RF Input Return Loss 8 48 Module Veri cation Tests ...
Страница 257: ...Test 25 LO Input Return Loss Figure 8 27 Equipment Setup for Test 25 LO Input Return Loss 8 52 Module Veri cation Tests ...
Страница 267: ......
Страница 271: ...Front Panel Figure 9 2 Front Panel Removal Replacement 1 of 2 9 4 Replacing Major Assemblies ...
Страница 272: ...Front Panel Figure 9 3 Front Panel Removal Replacement 2 of 2 Replacing Major Assemblies 9 5 ...
Страница 274: ...Rear Panel Figure 9 4 Rear Panel Removal Replacement 1 or 2 Replacing Major Assemblies 9 7 ...
Страница 275: ...Rear Panel Figure 9 5 Rear Panel Removal Replacement 2 or 2 9 8 Replacing Major Assemblies ...
Страница 277: ...Bandpass Filter Figure 9 6 Bandpass Filter Removal Replacement 1 or 2 9 10 Replacing Major Assemblies ...
Страница 278: ...Bandpass Filter Figure 9 7 Bandpass Filter Removal Replacement 2 or 2 Replacing Major Assemblies 9 11 ...
Страница 280: ...A1 Leveling Ampli er Figure 9 8 A1 Leveling Ampli er Removal Replacement Replacing Major Assemblies 9 13 ...
Страница 282: ...A2 LO Multiplier Figure 9 9 A2 LO Multiplier Removal Replacement Replacing Major Assemblies 9 15 ...
Страница 284: ...A3 SYTFMD Figure 9 10 A3 SYTFMD Removal Replacement Replacing Major Assemblies 9 17 ...
Страница 286: ...A4 Input Attenuator Figure 9 11 A4 Input Attenuator Removal Replacement 1 of 2 Replacing Major Assemblies 9 19 ...
Страница 287: ...A4 Input Attenuator Figure 9 12 A4 Input Attenuator Removal Replacement 2 of 2 9 20 Replacing Major Assemblies ...
Страница 289: ...A5 VCO Sampler Figure 9 13 A5 VCO Sampler Removal Replacement 1 of 3 9 22 Replacing Major Assemblies ...
Страница 290: ...A5 VCO Sampler Figure 9 14 A5 VCO Sampler Removal Replacement 2 of 3 Replacing Major Assemblies 9 23 ...
Страница 291: ...A5 VCO Sampler Figure 9 15 A5 VCO Sampler Removal Replacement 3 of 3 9 24 Replacing Major Assemblies ...
Страница 294: ...A7 2nd Mixer Figure 9 17 A7 2nd Mixer Removal Replacement Replacing Major Assemblies 9 27 ...
Страница 298: ...A9 2nd Converter PLL Figure 9 19 A9 2nd Converter PLL Removal Replacement 1 of 2 Replacing Major Assemblies 9 31 ...
Страница 299: ...A9 2nd Converter PLL Figure 9 20 A9 2nd Converter PLL Removal Replacement 2 of 2 9 32 Replacing Major Assemblies ...
Страница 301: ...A10 Power Supply Driver Figure 9 21 A10 Power Supply Driver Removal Replacement 9 34 Replacing Major Assemblies ...
Страница 303: ...A11 Last Converter Figure 9 22 A11 Last Converter Removal Replacement 1 of 2 9 36 Replacing Major Assemblies ...
Страница 304: ...A11 Last Converter Figure 9 23 A11 Last Converter Removal Replacement 2 of 2 Replacing Major Assemblies 9 37 ...
Страница 309: ...A15 RF 1st Converter Figure 9 27 A15 RF 1st Converter Removal Replacement 9 42 Replacing Major Assemblies ...
Страница 311: ...A16 Low Pass Filter Figure 9 28 A16 Low Pass Filter Removal Replacement 9 44 Replacing Major Assemblies ...
Страница 313: ...A17 Low Pass Filter Figure 9 29 A17 Low Pass Filter Removal Replacement 9 46 Replacing Major Assemblies ...
Страница 329: ......