T
est
03.
Attenuator
A
ccuracy
The
purpose
of
this
verication
test
is
to
measure
the
DUT
input
attenuator
switching
error
,
referenced
to
the
10
dB
input
attenuator
setting.
This
is
a
nal
test.
1.
Connect
the
equipment
as
shown
in
the
test
setup
illustration
for
this
procedure
.
2.
Set
the
mainframe
line
switch
to
ON.
3.
Load
and
run
T
est
03.
Attenuator
A
ccuracy
.
Refer
to
Chapter
2
for
information
about
running
the
software
.
When
performing
this
verication
test,
the
level
generator
provides
21.4
MHz,
050
dBm,
at
the
DUT
RF
INPUT
.
The
calibrated
spectrum
analyzer
connects
to
the
DUT
21.4
MHz
OUT
.
The
local
oscillator
source
tunes
to
center
the
RF
INPUT
signal.
The
DUT
input
attenuation
is
10
dB,
and
the
frequency
band
is
1H0.
The
calibrated
spectrum
analyzer
measures
the
21.4
MHz
OUT
as
a
reference
,
and
then
the
attenuator
steps
from
0
to
70
dB
in
10
dB
steps
.
At
each
step
the
level
generator
power
level
increases
10
dB
to
maintain
a
constant
amplitude
at
the
21.4
MHz
OUT
.
The
calibrated
spectrum
analyzer
measures
each
step
amplitude
,
and
then
subtracts
the
reference
measurement.
The
test
compares
the
dierence
with
amplitude
accuracy
test
limits
.
Module
V
erication
T
ests
8-9
Содержание 70908A
Страница 7: ...Servicing at a Glance vi ...
Страница 19: ......
Страница 44: ...Figure 2 1 Main Menu Softkeys Module Veri cation Software 2 19 ...
Страница 45: ...Figure 2 2 Mass Storage Menu and Parameter Menu Softkeys 2 20 Module Veri cation Software ...
Страница 46: ...Figure 2 3 Equipment Menu and HP MSIB Map Screen Menu Softkeys Module Veri cation Software 2 21 ...
Страница 47: ...Figure 2 4 Test Menu Softkeys 2 22 Module Veri cation Software ...
Страница 55: ......
Страница 61: ......
Страница 88: ...Troubleshooting the A1 Leveling Ampli er Figure 4 2 Notch Filter Waveform Troubleshooting 4 27 ...
Страница 94: ...Troubleshooting the A3 SYTFMD Figure 4 5 Discriminator Output Waveform Troubleshooting 4 33 ...
Страница 117: ...Overall Block Diagram of RF Section Back page for Foldout Goes Here This is just a place holder 4 56 Troubleshooting ...
Страница 125: ......
Страница 128: ...External Frequency Reference Figure 6 1 Preferred Frequency Reference Connections System Calibration 6 3 ...
Страница 130: ...External Frequency Reference Figure 6 4 Using an HP 8566B Spectrum Analyzer and a House Standard System Calibration 6 5 ...
Страница 139: ......
Страница 205: ......
Страница 209: ...Test 01 Flatness Figure 8 2 Equipment Setup for Test 01 Flatness 8 4 Module Veri cation Tests ...
Страница 253: ...Test 23 RF Input Return Loss Figure 8 25 Equipment Setup for Test 23 RF Input Return Loss 8 48 Module Veri cation Tests ...
Страница 257: ...Test 25 LO Input Return Loss Figure 8 27 Equipment Setup for Test 25 LO Input Return Loss 8 52 Module Veri cation Tests ...
Страница 267: ......
Страница 271: ...Front Panel Figure 9 2 Front Panel Removal Replacement 1 of 2 9 4 Replacing Major Assemblies ...
Страница 272: ...Front Panel Figure 9 3 Front Panel Removal Replacement 2 of 2 Replacing Major Assemblies 9 5 ...
Страница 274: ...Rear Panel Figure 9 4 Rear Panel Removal Replacement 1 or 2 Replacing Major Assemblies 9 7 ...
Страница 275: ...Rear Panel Figure 9 5 Rear Panel Removal Replacement 2 or 2 9 8 Replacing Major Assemblies ...
Страница 277: ...Bandpass Filter Figure 9 6 Bandpass Filter Removal Replacement 1 or 2 9 10 Replacing Major Assemblies ...
Страница 278: ...Bandpass Filter Figure 9 7 Bandpass Filter Removal Replacement 2 or 2 Replacing Major Assemblies 9 11 ...
Страница 280: ...A1 Leveling Ampli er Figure 9 8 A1 Leveling Ampli er Removal Replacement Replacing Major Assemblies 9 13 ...
Страница 282: ...A2 LO Multiplier Figure 9 9 A2 LO Multiplier Removal Replacement Replacing Major Assemblies 9 15 ...
Страница 284: ...A3 SYTFMD Figure 9 10 A3 SYTFMD Removal Replacement Replacing Major Assemblies 9 17 ...
Страница 286: ...A4 Input Attenuator Figure 9 11 A4 Input Attenuator Removal Replacement 1 of 2 Replacing Major Assemblies 9 19 ...
Страница 287: ...A4 Input Attenuator Figure 9 12 A4 Input Attenuator Removal Replacement 2 of 2 9 20 Replacing Major Assemblies ...
Страница 289: ...A5 VCO Sampler Figure 9 13 A5 VCO Sampler Removal Replacement 1 of 3 9 22 Replacing Major Assemblies ...
Страница 290: ...A5 VCO Sampler Figure 9 14 A5 VCO Sampler Removal Replacement 2 of 3 Replacing Major Assemblies 9 23 ...
Страница 291: ...A5 VCO Sampler Figure 9 15 A5 VCO Sampler Removal Replacement 3 of 3 9 24 Replacing Major Assemblies ...
Страница 294: ...A7 2nd Mixer Figure 9 17 A7 2nd Mixer Removal Replacement Replacing Major Assemblies 9 27 ...
Страница 298: ...A9 2nd Converter PLL Figure 9 19 A9 2nd Converter PLL Removal Replacement 1 of 2 Replacing Major Assemblies 9 31 ...
Страница 299: ...A9 2nd Converter PLL Figure 9 20 A9 2nd Converter PLL Removal Replacement 2 of 2 9 32 Replacing Major Assemblies ...
Страница 301: ...A10 Power Supply Driver Figure 9 21 A10 Power Supply Driver Removal Replacement 9 34 Replacing Major Assemblies ...
Страница 303: ...A11 Last Converter Figure 9 22 A11 Last Converter Removal Replacement 1 of 2 9 36 Replacing Major Assemblies ...
Страница 304: ...A11 Last Converter Figure 9 23 A11 Last Converter Removal Replacement 2 of 2 Replacing Major Assemblies 9 37 ...
Страница 309: ...A15 RF 1st Converter Figure 9 27 A15 RF 1st Converter Removal Replacement 9 42 Replacing Major Assemblies ...
Страница 311: ...A16 Low Pass Filter Figure 9 28 A16 Low Pass Filter Removal Replacement 9 44 Replacing Major Assemblies ...
Страница 313: ...A17 Low Pass Filter Figure 9 29 A17 Low Pass Filter Removal Replacement 9 46 Replacing Major Assemblies ...
Страница 329: ......