10
MHz
and
100
MHz
Output
P
ower
T
est
Figure
4-2.
10
MHz
and
100
MHz
Output
P
ower
T
est
Setup
Procedure
1.
Calibrate
the
power
meter
and
sensor
according
to
instructions
in
the
power
meter's
Operating
Manual.
2.
Set
the
synthesized-source
frequency
to
10.0
MHz
and
amplitude
to
0
dBm.
3.
Connect
the
power
sensor
to
the
DUT's
10
MHz
output.
The
power-meter
reading
should
be
0
dBm
02
dB
,
+3
dB
(02
dBm
to
+3
dBm).
4.
Connect
the
power
sensor
to
the
DUT's
100
MHz
output.
The
power-meter
reading
should
be
0
dBm
02
dB
,
+3
dB
(02
dBm
to
+3
dBm).