Checking
Module
Operation
The
HP
70310A
Precision
Frequency
Reference
can
be
congured
as
a
separate
frequency
reference
(installed
in
an
HP
70000
Modular
Measurement
System
mainframe)
or
as
a
slave
in
an
HP
70000
Modular
Measurement
System.
Module-operation
checks
are
given
below
for
the
following
congurations:
HP
70310A
module
used
as
a
separate
frequency
reference
.
HP
70310A
module
used
as
a
slave
in
an
HP
70000
Modular
Measurement
System
that
has
an
HP
70900A
Local
Oscillator
as
the
system
master
.
HP
70310A
as
a
Separate
Frequency
Reference
The
operation
of
the
HP
70310A
module
as
a
separate
frequency
reference
is
veried
by
checking
the
results
of
the
module's
power-on
self-test.
The
results
of
the
self-test
can
be
determined
by
examining
the
status
of
the
module's
LEDs
.
An
HP
70000
Modular
Measurement
System
display
can
also
be
used
to
check
for
error
messages
.
Front-P
anel
LEDs
The
power-on
self-test
runs
automatically
when
power
is
rst
applied
to
the
module
.
The
listing
below
describes
what
the
status
of
the
LEDs
should
be
immediately
after
the
self-test
has
run.
The
A
CT
LED
should
be
o.
The
ERR
LED
should
be
on
if
the
oven/oscillator
is
still
cold.
The
ERR
LED
may
stay
on
when
the
O
VEN
COLD
LED
goes
o.
If
this
happens
,
cycling
the
power
(turning
the
mainframe
o,
then
quickly
turning
it
on
again)
should
result
in
the
ERR
LED
going
o.
HP
70310A
Option
002
modules
should
not
have
the
ERR
LED
go
on.
The
O
VEN
COLD
LED
should
be
on
if
the
oven/oscillator
is
still
cold.
This
LED
should
go
o
approximately
fteen
minutes
after
power-on
if
the
module
was
stored
at
room
temperature
.
HP
70310A
Option
002
modules
should
not
have
the
O
VEN
COLD
LED
go
on.
The
EXTERNAL
REF
LED
should
be
o
if
the
internal
frequency
reference
is
being
used.
The
EXTERNAL
REF
LED
should
go
on
when
an
appropriate
external
reference
is
connected
to
the
HP
70310A
module
.
The
PLL
ERROR
LED
should
be
o.
Checking
for
Error
Messages
If
the
HP
70310A
module's
ERR
LED
remains
on,
an
HP
70000
Modular
Measurement
System
display
can
be
used
to
determine
what
error
is
occurring.
Note
Remove
all
master
elements
or
systems
from
the
HP-MSIB
before
using
the
following
troubleshooting
procedure
.
If
there
are
any
master
elements
or
systems
on
the
HP-MSIB
when
this
procedure
is
performed,
the
system
or
master
may
cease
communication
and
the
procedure
may
not
work.
Installation
2-11