10
MHz
Output
Spurs
T
est
5.
Connect
the
RF
OUTPUT
of
the
synthesized
source
to
the
DUT's
EXT
REF
input.
Press
4
RECALL
5
4
1
5
on
the
synthesized-source
front
panel
to
set
the
synthesized-source
frequency
and
amplitude
to
the
settings
saved
in
step
1.
6.
Set
the
spectrum-analyzer
center
frequency
to
10
MHz,
frequency
span
to
50
kHz,
and
reference
level
to
+10
dBm.
7.
Set
the
spectrum-analyzer
marker
to
the
highest
peak
and
note
the
carrier
amplitude
.
It
should
be
approximately
0
dBm.
8.
Set
the
spectrum-analyzer
center
frequency
to
1
MHz,
frequency
span
to
50
kHz,
and
reference
level
to
020
dBm .
All
signals
should
be
at
least
80
dB
below
the
carrier
level
noted
in
step
7.
9.
Change
the
spectrum-analyzer
center
frequency
setting
to
9
MHz.
All
signals
should
still
be
at
least
80
dB
below
the
carrier
level
noted
in
step
7.
10.
Change
the
spectrum-analyzer
center
frequency
setting
to
11
MHz.
All
signals
should
still
be
at
least
80
dB
below
the
carrier
level
noted
in
step
7.
11.
Press
4
RECALL
5
4
2
5
on
the
synthesized-source
front
panel
to
set
the
synthesized-source
frequency
and
amplitude
to
the
settings
saved
in
step
2.
12.
Set
the
spectrum-analyzer
center
frequency
to
2
MHz,
frequency
span
to
50
kHz,
and
reference
level
to
020
dBm .
All
signals
should
be
at
least
80
dB
below
the
carrier
level
noted
in
step
7.
13.
Change
the
spectrum-analyzer
center
frequency
setting
to
8
MHz.
All
signals
should
still
be
at
least
80
dB
below
the
carrier
level
noted
in
step
7.
14.
Change
the
spectrum-analyzer
center
frequency
setting
to
12
MHz.
All
signals
should
still
be
at
least
80
dB
below
the
carrier
level
noted
in
step
7.
15.
Press
4
RECALL
5
4
3
5
on
the
synthesized-source
front
panel
to
set
the
synthesized-source
frequency
and
amplitude
to
the
settings
saved
in
step
3.
16.
Set
the
spectrum-analyzer
center
frequency
to
5
MHz,
frequency
span
to
50
kHz,
and
reference
level
to
020
dBm .
All
signals
should
be
at
least
80
dB
below
the
carrier
level
noted
in
step
7.
17.
Change
the
spectrum-analyzer
center
frequency
to
15
MHz.
All
signals
should
still
be
at
least
80
dB
below
the
carrier
level
noted
in
step
7.
18.
Press
4
RECALL
5
4
4
5
on
the
synthesized-source
front
panel
to
set
the
synthesized-source
frequency
and
amplitude
to
the
settings
saved
in
step
4.
19.
Set
the
spectrum-analyzer
center
frequency
to
10
MHz,
frequency
span
to
5
kHz,
RES
BW
to
30
Hz,
VIDEO
BW
to
10
Hz,
and
reference
level
to
+10
dBm.
20.
Set
the
spectrum-analyzer
marker
to
the
highest
peak
and
note
the
carrier
amplitude
.
It
should
be
approximately
0
dBm.
21.
Set
the
spectrum-analyzer
center
frequency
to
10.04
MHz,
frequency
span
to
5
kHz,
RES
BW
to
30
Hz,
VIDEO
BW
to
10
Hz,
and
reference
level
to
020
dBm.
All
signals
should
be
at
least
80
dB
below
the
carrier
level
noted
in
step
20.
22.
Change
the
spectrum-analyzer
center
frequency
setting
to
10.08
MHz,
frequency
span
to
10
kHz,
and
VIDEO
BW
to
30
Hz.
All
signals
should
be
at
least
80
dB
below
the
carrier
level
noted
in
step
20.
V
erication
4-9
Содержание 70310A
Страница 6: ......
Страница 14: ...Figure 1 2 HP 70310A Module with Accessories 1 4 General Information ...
Страница 30: ...Figure 2 3 External Power Pack Installation 2 4 Installation ...
Страница 64: ......
Страница 66: ......