Chapter 2 Operating Your Universal Counter
Using the Utility Menu
Operating Guide
2-61
2
Example Procedure for Running the Self Test
1
Press and hold Recall (Utility) key, then cycle POWER key.
2
Press Recall (Utility) key until
TEST:
is displayed.
3
Press any one of the arrow keys to select the desired menu choice in the
menu item (for example, the
ALL?
menu choice).
4
Press Enter key to start the test.
5
If you are asked to do something, press Enter key when you have
completed the requested task.
A pass or fail message will be displayed.
To Configure the RS-232 Serial Port for Printing
Software pacing, baud rate, parity, data bits, and stop bits parameters for
each port are user-selectable and independent of configuration of the
other.
The table below lists the configuration factory-default values for the
RS-232 port.
1
When parity is EVEN or ODD, data bits are fixed at 7. When parity is NONE, data bits are fixed at 8.
Parameter
Default
Possible Choices
Software Pacing
XON
XON or NONE
Baud Rate
9600
300, 1200, 2400, 9600, or 19200
Parity
NONE
EVEN or ODD
1
NONE
Data Bits
8
FIXED at 7
FIXED at 8
Stop Bits
1
FIXED
Full Duplex
ON
ON or OFF
DTR
HIGH
PACE, HIGH, LIMIT
Содержание 53131A
Страница 1: ...Operating Guide HP 53131A 132A 225 MHz Universal Counter ...
Страница 2: ......
Страница 26: ...xxiv Operating Guide ...
Страница 41: ...1 Getting Started ...
Страница 77: ...2 Operating Your Universal Counter Operator s Reference ...
Страница 171: ...3 Specifications ...
Страница 196: ... Manual Part Number 53131 90055 Printed in U S A MAY 1999 ...