Chapter 2 Operating Your Universal Counter
Using the Utility Menu
2-62
Operating Guide
2
Configuring the RS-232 serial port for printing requires that you:
•
set the hardware pacing in the DTR (Data Terminal Ready) menu item,
•
turn PRINT:ON in the SAVE&PRINT menu,
•
set the baud rate in the BAUD menu item,
•
set the parity type in the PARITY menu item, and
•
set the software pacing in the SW PACE menu item.
Setting the Hardware Pacing
1
Press and hold Recall (Utility) key, then cycle POWER key.
2
Press Recall (Utility) key until
DTR:
is displayed.
3
Press any one of the arrow keys until the desired hardware pacing is
displayed.
NOTE
•
Choose
HW PACE
to use the DTR line for pacing.(
CAUTION:
You
should only choose this option if your printer and cable
support DTR handshaking.)
•
Choose
HIGH
to disable hardware pacing.
The Counter’s talk-only RS-232 port is now set to be the driver for a
RS-232 serial printer. The hardware pacing choice is stored in non-volatile
memory, and does not change when power has been off or after a remote
interface reset.
Setting the Baud Rate
You can select one of five baud rates for RS-232 operation. The rate is set
to 9600 baud when the Counter is shipped from the factory.
1
Press Recall (Utility) key until
BAUD:
is displayed.
2
Press any one of the arrow keys until the desired baud rate is displayed.
The baud rate is stored in non-volatile memory, and does not change when
power has been off or after a remote interface reset.
Содержание 53131A
Страница 1: ...Operating Guide HP 53131A 132A 225 MHz Universal Counter ...
Страница 2: ......
Страница 26: ...xxiv Operating Guide ...
Страница 41: ...1 Getting Started ...
Страница 77: ...2 Operating Your Universal Counter Operator s Reference ...
Страница 171: ...3 Specifications ...
Страница 196: ... Manual Part Number 53131 90055 Printed in U S A MAY 1999 ...