Chapter 2 Operating Your Universal Counter
Using the Save and Recall Menus
Operating Guide
2-53
2
To Use the Save Function
The following step-by-step procedure describes how to SAVE a simple
measurement setup to quickly get you familiar with the Save function.
You will SAVE the following in register 1 (SAVE: 1):
•
the Period measurement function,
•
the input conditioning parameters:
–
auto trigger to off
–
trigger level to 1.5V
–
input impedance to 50
Ω
–
input coupling to DC
1
Cycle POWER key to preset the Counter setup.
The Counter is now ready to measure the frequency of a signal applied to
CHANNEL 1 input as indicated by the Freq and Ch1 annunciators.
2
Press Other Meas key until
PERIOD 1
is displayed.
The Counter is now ready to measure the period of a signal applied to
CHANNEL 1 input as indicated by the Period and Ch1 annunciators.
3
Press Trigger/Sensitivity key.
AUTO TRG: ON
is displayed.
4
Press any one of the arrow keys.
AUTO TRG: OFF
is displayed.
5
Press Trigger/Sensitivity key, or Enter key.
LEVEL: 0.000V
is displayed.
6
Using the appropriate arrow keys, set the trigger level to 1.5V.