Turn the control dial to RUN.
Piggyback delivery begins.
When secondary delivery
completes, Plum XLs with
software revisions 1.02 and
earlier beep five times and begin
primary delivery.
(Optional): Set the panel lockout
switch to locked to prevent
unauthorized tampering of the
device (see
Section 6.8, Lockout
).
6.4
Backpriming
Backpriming is typically used to clear accumulated air
from the cassette or to clear air from the secondary line
without disconnecting the administration set from the
patient. AIR IN LINE and BACKPRIMING display on the
screen and an alarm sounds when air is detected in the
cassette. Fluid is backprimed from the primary line up
through the secondary inlet port to expel the air.
To expel air from the cassette when using primary
delivery (secondary inlet port is capped or resealed), or
when using piggyback delivery, proceed as follows:
Turn the control dial to
HOLD/RESET (the alarm is
silenced).
OFF
CHARGE
SET RATE
SET VTBI
RUN
HOLD/RESET
CLEAR
VOL