System
Operating
Manual
31
LOW BATTERY
Approximately 30
minutes of battery
power remains
Connect to AC power
Note:
Pressing the
[SILENCE] key mutes
the audible alarm for 15
minutes from the time
the LOW BATTERY
alarm occurred
Note:
When the battery
discharges, pumping
stops and the alarm
sounds continuously for
one minute before the
device shuts down
completely
OCCLUSION
Clamp closed
Tubing kinked
Possible clotted
catheter
Open clamps
Unkink tubing
Check IV site
TURN TO RUN
Control dial is not in
OFF CHARGE or
RUN setting and no
key is pressed for
five minutes
Turn control dial to
RUN, OFF CHARGE, or
HOLD/RESET
MESSAGE
POSSIBL E CA USE
CORRECTIVE ACTION