6
Plum
XL
Micro/Macro
After the cassette door is closed, turn the control dial to
SET RATE.
The LCD screen displays all the
symbols briefly. Verify that the
screen display exactly matches
the illustration shown at the left.
If the LCD screen does not match
the illustration, remove the
Plum XL from service and contact
the hospital repair facility or
Hospira Technical Support
Operations.
After the general LCD self-test,
the LCD screen displays
duplicated segments of the
numeric display. Verify that the
screen display alternates exactly
as shown in the illustrations at
the left. If the display does not
match the illustrations, remove
the Plum XL from service and
contact the hospital repair facility
or Hospira Technical Support
Operations.
After the Plum XL completes self-testing, disconnect it
from AC power and confirm that BATTERY displays on
the screen (indicating battery power is in use). To ensure
the battery is fully charged, remove the administration
set, then reconnect the Plum XL to AC power for a
minimum of eight hours in the OFF CHARGE setting.
If an alarm occurs during the self-test, note the message,
then take the appropriate corrective action (see
Section
7.0, TROUBLESHOOTING
). Repeat the self-test. If the
alarm recurs, remove the Plum XL from service and
contact Hospira Technical Support Operations.
CAUTION: Do not operate the Plum XL with the battery
removed. The use of a properly maintained and charged
battery ensures proper operation. As always in the event
of an AC power interruption or failure verify infusion
pum p settings
Содержание Micro Macro plum XL
Страница 6: ...i iv v P Pl lu um m X XL L M Mi ic cr ro o M Ma ac cr ro o NOTES NOTES...
Страница 54: ...48 Plum XL Micro Macro NOTES...
Страница 57: ...System Operating Manual 51 NOTES...