28
Plum
XL
Micro/Macro
6.6
Titration
Titration is the incremental adjustment of the fluid
delivery rate while pumping (primary or secondary) is in
progress.
To titrate fluid delivery, hold down the [TITRATE] key
and press the
or
key to increase or decrease the
delivery rate.
6.7
Clear Volume
CLEAR VOL erases the total volume delivered from
memory.
The total volume delivered is the total amount of
fluid, both primary and secondary, delivered to the
patient.
To clear the total volume, turn the
control dial to CLEAR VOL. Four
beeps sound before the total
volume is cleared.
To cancel the clear total
volume function, turn the control
dial away from the CLEAR VOL
setting before the fourth beep
sounds (e.g., turn the control dial
to HOLD/RESET).
6.8
Lockout
To avoid unauthorized tampering of the Plum XL, set the
panel LOCKOUT switch to on. When the lockout switch
is on, the word LOCKED appears on the LCD screen and
the keys are inactive. If the control dial is moved, the
LOCKED alarm will sound. To clear the alarm, set the
LOCKOUT switch to off.
OFF
CHARGE
SET RATE
SET VTBI
RUN
HOLD/RESET
CLEAR
VOL
Note:
Note: